Preconcentration and separation of double-stranded DNA fragments by electrophoresis in plastic microfluidic devices
- PMID: 14613206
- DOI: 10.1002/elps.200305594
Preconcentration and separation of double-stranded DNA fragments by electrophoresis in plastic microfluidic devices
Abstract
We have evaluated double-stranded DNA separations in microfluidic devices which were designed to couple a sample preconcentration step based on isotachophoresis (ITP) with a zone electrophoretic (ZE) separation step as a method to increase the concentration limit of detection in microfluidic devices. Developed at ACLARA BioSciences, these LabCard trade mark devices are plastic 32 channel chips, designed with a long sample injection channel segment to increase the sample loading. These chips were designed to allow stacking of the sample into a narrow band using discontinuous ITP buffers, and subsequent separation in the ZE mode in sieving polymer solutions. Compared to chip ZE, the sensitivity was increased by 40-fold and we showed baseline resolution of all fragments in the PhiX174/HaeIII DNA digest. The total analysis time was 3 min/sample, or less than 100 min per LabCard device. The resolution for multiplexed PCR samples was the same as obtained in chip ZE. The limit of detection was 9 fg/microL of DNA in 0.1xpolymerase chain reaction (PCR) buffers using confocal fluorescence detection following 488 nm laser excitation with thiazole orange as the fluorescent intercalating dye.
Similar articles
-
Isotachophoresis preconcentration integrated microfluidic chip for highly sensitive genotyping of the hepatitis B virus.J Chromatogr B Analyt Technol Biomed Life Sci. 2006 Nov 21;844(1):32-8. doi: 10.1016/j.jchromb.2006.06.041. Epub 2006 Aug 8. J Chromatogr B Analyt Technol Biomed Life Sci. 2006. PMID: 16899416
-
Sample pre-concentration by isotachophoresis in microfluidic devices.J Chromatogr A. 2002 Dec 6;979(1-2):69-80. doi: 10.1016/s0021-9673(02)01246-3. J Chromatogr A. 2002. PMID: 12498234
-
High-sensitive analysis of DNA fragments by capillary gel electrophoresis using transient isotachophoresis preconcentration and fluorescence detection.J Chromatogr A. 2009 Apr 24;1216(17):3602-5. doi: 10.1016/j.chroma.2009.01.066. Epub 2009 Jan 29. J Chromatogr A. 2009. PMID: 19211105
-
Sample introduction techniques for microfabricated separation devices.Electrophoresis. 2004 Jan;25(2):229-42. doi: 10.1002/elps.200305742. Electrophoresis. 2004. PMID: 14743476 Review.
-
Miniaturizing free-flow electrophoresis - a critical review.Electrophoresis. 2008 Mar;29(5):977-93. doi: 10.1002/elps.200700725. Electrophoresis. 2008. PMID: 18232029 Review.
Cited by
-
Optofluidic in situ maskless lithography of charge selective nanoporous hydrogel for DNA preconcentration.Biomicrofluidics. 2010 Dec 30;4(4):43014. doi: 10.1063/1.3516037. Biomicrofluidics. 2010. PMID: 21267091 Free PMC article.
-
Comparing polyelectrolyte multilayer-coated PMMA microfluidic devices and glass microchips for electrophoretic separations.Electrophoresis. 2009 Dec;30(24):4245-50. doi: 10.1002/elps.200900403. Electrophoresis. 2009. PMID: 20013912 Free PMC article.
-
A parallel microfluidic channel fixture fabricated using laser ablated plastic laminates for electrochemical and chemiluminescent biodetection of DNA.Biomicrofluidics. 2011 Dec;5(4):44115-4411514. doi: 10.1063/1.3664694. Epub 2011 Dec 15. Biomicrofluidics. 2011. PMID: 22276087 Free PMC article.
-
On-chip DNA preconcentration in different media conductivities by electrodeless dielectrophoresis.Biomicrofluidics. 2015 Sep 30;9(5):054115. doi: 10.1063/1.4932177. eCollection 2015 Sep. Biomicrofluidics. 2015. PMID: 26487901 Free PMC article.
-
An Automated and Miniaturized Rotating-Disk Device for Rapid Nucleic Acid Extraction.Micromachines (Basel). 2019 Mar 22;10(3):204. doi: 10.3390/mi10030204. Micromachines (Basel). 2019. PMID: 30909465 Free PMC article.
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Other Literature Sources