Skip to main page content
U.S. flag

An official website of the United States government

Dot gov

The .gov means it’s official.
Federal government websites often end in .gov or .mil. Before sharing sensitive information, make sure you’re on a federal government site.

Https

The site is secure.
The https:// ensures that you are connecting to the official website and that any information you provide is encrypted and transmitted securely.

Access keys NCBI Homepage MyNCBI Homepage Main Content Main Navigation
Comparative Study
. 2003 Dec;36(12):1785-96.
doi: 10.1016/s0021-9290(03)00231-8.

Cartilage interstitial fluid load support in unconfined compression

Affiliations
Comparative Study

Cartilage interstitial fluid load support in unconfined compression

Seonghun Park et al. J Biomech. 2003 Dec.

Abstract

Under physiological conditions of loading, articular cartilage is subjected to both compressive strains, normal to the articular surface, and tensile strains, tangential to the articular surface. Previous studies have shown that articular cartilage exhibits a much higher modulus in tension than in compression, and theoretical analyses have suggested that this tension-compression nonlinearity enhances the magnitude of interstitial fluid pressurization during loading in unconfined compression, above a theoretical threshold of 33% of the average applied stress. The first hypothesis of this experimental study is that the peak fluid load support in unconfined compression is significantly greater than the 33% theoretical limit predicted for porous permeable tissues modeled with equal moduli in tension and compression. The second hypothesis is that the peak fluid load support is higher at the articular surface side of the tissue samples than near the deep zone, because the disparity between the tensile and compressive moduli is greater at the surface zone. Ten human cartilage samples from six patellofemoral joints, and 10 bovine cartilage specimens from three calf patellofemoral joints were tested in unconfined compression. The peak fluid load support was measured at 79 +/- 11% and 69 +/- 15% at the articular surface and deep zone of human cartilage, respectively, and at 94 +/- 4% and 71 +/- 8% at the articular surface and deep zone of bovine calf cartilage, respectively. Statistical analyses confirmed both hypotheses of this study. These experimental results suggest that the tension-compression nonlinearity of cartilage is an essential functional property of the tissue which makes interstitial fluid pressurization the dominant mechanism of load support in articular cartilage.

PubMed Disclaimer

Figures

Figure 1
Figure 1
(a) Testing chamber for unconfined compression. The cylindrical cartilage sample (diameter b = 6 mm) is flush against a recessed free-draining porous filter (diameter a = 4.78 mm) communicating with a piezoresistive microchip pressure transducer. (b) The non-uniform pressure distribution at the interface of the cartilage and filter is averaged into a single time-varying value p0(t) measured by the pressure transducer over the footprint of the porous filter. In the annular region a/≤rb/2 the pressure is assumed to drop linearly to zero, producing a trapezoidal pressure profile p(r,t) along r.
Figure 2
Figure 2
(a) Typical response for the total load, W(t), and the load supported by interstitial fluid pressure, Wp(t), for a human Specimen in Experiment 1, with the articular surface facing the pressure transducer. Both W(t) and Wp(t) rise during the compressive phase of loading (0 ≤ t ≤ s) conducted at a constant strain rate (20% over 100 s), and reduce to zero during the unloading phase when the platen recedes at the same rate (100 ≤ t ≤ 200 s), because loading platen lift-off typically occurs during unloading. (b) Plot of Wp(t) versus W(t) during the loading phase (0 ≤ t ≤100 s), using the same data set as in (a). Linear regression on the data provides the slope dWp/dW which remains nearly constant during the entire loading phase.
Figure 3
Figure 3
(a) Typical response for W(t) and Wp(t), for a bovine Specimen in Experiment 2, with the articular surface facing the pressure transducer. Both W(t) and Wp(t) rise during the compressive phase of loading (0 ≤ t 500 s) conducted at a constant strain rate (10% over 500 s). For t > 500 s the applied deformation is maintained constant and W(t) relaxes to its equilibrium value while Wp(t) reduces to zero. (b) Plot of Wp(t) versus W(t) during the early portion of the loading phase (0 ≤ t ≤ 180 s), using the same data set as in (a). Linear regression on the linear portion of the data provides the slope dWp/dW and the load W0 where the compliance is approximately overcome.
Figure 4
Figure 4
(a) Fluid load support Wp/W' as a function of time (W'(t) = W(t) − W0) for the four tests of Experiment 2. (b) Theoretical fluid load support Wp/W predicted from the biphasic-CLE theory for the same loading protocol as in (a), using typical material properties HA =0.64 MPa, λ2=0.48 MPa, k=0.6×10−15 m4/N.s (Soltz and Ateshian, 2000b), and varying the ratio of the tensile to compressive moduli H+A − λ2/H−A − λ2 from 30 down to 1.
Figure 5
Figure 5
Testing configuration and results for finite element analyses of contact analysis of a spherical indenter on a cartilage layer (physiologic contact condition), unconfined compression and confined compression of a cylindrical cartilage sample, and indentation of a cartilage layer with a flat porous permeable probe. Results include (a) the interstitial fluid pressurization, with arrows indicating the relative interstitial fluid flux; (b) the minimum principal effective stress, with lines and arrows indicating their principal direction; and (c) the maximum principal effective stress, with lines and arrows indicating their principal directions.
Figure 6
Figure 6
Theoretical analysis of confined compression of a biphasic cylindrical cartilage plug, taking into account pressure transducer impedance. (a) Schematic of loading configuration. (b) Theoretical prediction of interstitial fluid load support for various ratios of the pressure transducer sensor and cartilage compressive aggregate moduli (HAt/HA).

Similar articles

Cited by

References

    1. Akizuki S, Mow VC, Muller F, Pita JC, Howell DS, Manicourt DH. Tensile properties of human knee joint cartilage: I. Influence of ionic conditions, weight bearing, and fibrillation on the tensile modulus. J Orthop Res. 1986;4:379–392. - PubMed
    1. Armstrong CG, Mow VC. Variations in the intrinsic mechanical properties of human articular cartilage with age, degeneration, and water content. J Bone Joint Surg Am. 1982;64:88–94. - PubMed
    1. Armstrong CG, Lai WM, Mow VC. An analysis of the unconfined compression of articular cartilage. J Biomech Eng. 1984;106:165–173. - PubMed
    1. Ateshian GA. A theoretical formulation for boundary friction in articular cartilage. J Biomech Eng. 1997;119:81–86. ASME. - PubMed
    1. Ateshian GA, Lai WM, Zhu WB, Mow VC. An asymptotic solution for the contact of two biphasic cartilage layers. J Biomech. 1994;27:1347–1360. - PubMed

LinkOut - more resources