Comparison of indirect CsI/a:Si and direct a:Se digital radiography. An assessment of contrast and detail visualization
- PMID: 14616206
- DOI: 10.1080/02841850312331287769
Comparison of indirect CsI/a:Si and direct a:Se digital radiography. An assessment of contrast and detail visualization
Abstract
Purpose: To assess and quantify the image quality at two dose levels for an amorphous Silicon (a:Si) Cesium Iodide (CsI) flat panel system compared with a direct amorphous Selenium (a:Se) digital radiography system.
Material and methods: A contrast detectability test was performed employing the CDRAD-phantom at mAs-values leading to approximately equal phantom entrance doses of 41.4, 57.9, 75.1 and 120.8 micro Gy for the a:Se and 39.9, 58.4, 75.6 and 117.9 micro Gy for the CsI system. Images were presented to 4 independent observers. For quantitative comparison, the image quality figure (IQF) was calculated. Statistical analysis was performed using Pearson's correlation and the Wilcoxon test. A ROC-analysis was performed employing the TRG-phantom in a high- and a low-dose setting leading to entrance doses of 126.2 and 35 micro Gy for the direct, and 125.9 and 34.4 micro Gy for the indirect system. Statistical significance was evaluated using the Wilcoxon test.
Results: The flat panel a:Si digital system provided superior results compared with the a:Se drum digital system with respect to low-dose settings for CDRAD-phantom and ROC-analysis, ensuring a better image quality with respect to contrast and detail detectability. Higher-dose settings provided similar results for both systems.
Conclusion: Image quality of a:Si flat panel digital radiography proved to be superior to a:Se drum digital radiography using low-dose settings. If the primary target is dose reduction indirect flat panel technology should be used.
Similar articles
-
Flat panel digital radiography compared with storage phosphor computed radiography: assessment of dose versus image quality in phantom studies.Invest Radiol. 2002 Nov;37(11):609-14. doi: 10.1097/00004424-200211000-00004. Invest Radiol. 2002. PMID: 12393973
-
Amorphous silicon, flat-panel, x-ray detector versus storage phosphor-based computed radiography: contrast-detail phantom study at different tube voltages and detector entrance doses.Invest Radiol. 2003 Apr;38(4):212-20. doi: 10.1097/01.RLI.0000057032.41715.15. Invest Radiol. 2003. PMID: 12649645
-
Contrast-detail phantom study for x-ray spectrum optimization regarding chest radiography using a cesium iodide-amorphous silicon flat-panel detector.Invest Radiol. 2004 Oct;39(10):610-8. doi: 10.1097/01.rli.0000138091.96320.f8. Invest Radiol. 2004. PMID: 15377940
-
Dose reduction in skeletal and chest radiography using a large-area flat-panel detector based on amorphous silicon and thallium-doped cesium iodide: technical background, basic image quality parameters, and review of the literature.Eur Radiol. 2004 May;14(5):827-34. doi: 10.1007/s00330-004-2243-2. Epub 2004 Feb 17. Eur Radiol. 2004. PMID: 14968260 Review.
-
Radiation dose and image quality in diagnostic radiology. Optimization of the dose-image quality relationship with clinical experience from scoliosis radiography, coronary intervention and a flat-panel digital detector.Acta Radiol Suppl. 2002 Mar;43(427):1-43. Acta Radiol Suppl. 2002. PMID: 12108231 Review.
Cited by
-
Single-exposure dual-energy subtraction chest radiography: detection of pulmonary nodules and masses in clinical practice.Eur Radiol. 2008 Jan;18(1):24-31. doi: 10.1007/s00330-007-0758-z. Epub 2007 Sep 27. Eur Radiol. 2008. PMID: 17899105
-
Investigation of Deconvolution Method with Adaptive Point Spread Function Based on Scintillator Thickness in Wavelet Domain.Bioengineering (Basel). 2024 Mar 28;11(4):330. doi: 10.3390/bioengineering11040330. Bioengineering (Basel). 2024. PMID: 38671752 Free PMC article.
-
Evaluation of chest and abdominal exposure dose appropriate for a digital image reader system incorporating a columnar-crystal structured phosphor plate and a contrast-detail phantom.Radiol Phys Technol. 2008 Jul;1(2):238-43. doi: 10.1007/s12194-008-0035-0. Epub 2008 Jul 5. Radiol Phys Technol. 2008. PMID: 20821154
-
Image quality and radiation dose in planar imaging - Image quality figure of merits from the CDRAD phantom.J Appl Clin Med Phys. 2019 Jul;20(7):151-159. doi: 10.1002/acm2.12649. Epub 2019 Jun 1. J Appl Clin Med Phys. 2019. PMID: 31152576 Free PMC article.
-
Introduction of a New Parameter for Evaluation of Digital Radiography System Performance.J Med Signals Sens. 2020 Jul 3;10(3):196-200. doi: 10.4103/jmss.JMSS_53_19. eCollection 2020 Jul-Sep. J Med Signals Sens. 2020. PMID: 33062611 Free PMC article.
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources