Serum cobalt levels after metal-on-metal total hip arthroplasty
- PMID: 14630848
- DOI: 10.2106/00004623-200311000-00017
Serum cobalt levels after metal-on-metal total hip arthroplasty
Abstract
Background: Systemic cobalt dissemination from the Metasul Co-28Cr-6Mo-0.2C metal-on-metal total hip prosthesis has been demonstrated in the first year after implantation. The aim of this prospective study was to monitor the serum cobalt concentrations in patients during the first five years after total hip arthroplasty with a metal-on-metal articulation.
Methods: A total hip arthroplasty was performed without cement in 100 consecutive patients who had either unilateral osteoarthritis or unilateral osteonecrosis. Fifty patients were randomized to be treated with a metal-on-metal articulation, and fifty patients, with a ceramic-on-polyethylene bearing. The femoral stem was made of a Ti-6Al-7Nb alloy, and the threaded acetabular cup was made of commercially pure titanium. Blood samples were taken before the operation and at multiple time-points for five years after the operation. Serum cobalt concentrations were measured with use of atomic absorption spectrometry.
Results: In the metal-on-metal group, the median serum cobalt concentration was 1 micro g/L at one year after surgery and 0.7 micro g/L at five years. The median of the serum cobalt concentrations measured from three to twelve months did not differ from the median of subsequent measurements, with the numbers available. The median serum cobalt level in the control group of patients treated with the ceramic-on-polyethylene articulation was below the detection limit at all time-points.
Conclusions: Systemic cobalt release from Metasul metal-on-metal articulations was demonstrated throughout the five-year study period. The median serum cobalt concentrations were found to be slightly above the detection limit and remained in a constant range. The serum cobalt concentrations did not reflect a so-called run-in wear period of the metal-on-metal articulations.
Similar articles
-
Serum-cobalt levels with metal-on-metal bearings in the cement-free total hip arthroplasty results covering two years; prospective study.Acta Chir Orthop Traumatol Cech. 2011;78(5):410-5. Acta Chir Orthop Traumatol Cech. 2011. PMID: 22094154 Clinical Trial.
-
Ten-year outcome of serum metal ion levels after primary total hip arthroplasty: a concise follow-up of a previous report*.J Bone Joint Surg Am. 2013 Mar 20;95(6):512-8. doi: 10.2106/JBJS.L.00471. J Bone Joint Surg Am. 2013. PMID: 23515985 Free PMC article. Clinical Trial.
-
Metal ion release from bearing wear and corrosion with 28 mm and large-diameter metal-on-metal bearing articulations: a follow-up study.J Bone Joint Surg Br. 2010 Jan;92(1):12-9. doi: 10.1302/0301-620X.92B1.22226. J Bone Joint Surg Br. 2010. PMID: 20044673 Clinical Trial.
-
Systemic toxicity related to metal hip prostheses.Clin Toxicol (Phila). 2014 Sep-Oct;52(8):837-47. doi: 10.3109/15563650.2014.944977. Epub 2014 Aug 16. Clin Toxicol (Phila). 2014. PMID: 25132471 Review.
-
The argument for the use of Metasul as an articulation surface in total hip replacement.Clin Orthop Relat Res. 2004 Dec;(429):80-5. doi: 10.1097/01.blo.0000150343.66755.79. Clin Orthop Relat Res. 2004. PMID: 15577470 Review.
Cited by
-
Peripheral Blood Lymphocyte Subpopulations in Patients Following Small Diameter Metal-On-Metal Total Hip Replacement at Long-Term Follow-Up.J Clin Med. 2020 Sep 6;9(9):2877. doi: 10.3390/jcm9092877. J Clin Med. 2020. PMID: 32899901 Free PMC article.
-
Immunologic adverse reaction associated with low-carbide metal-on-metal bearings in total hip arthroplasty.Clin Orthop Relat Res. 2010 Aug;468(8):2135-42. doi: 10.1007/s11999-009-1187-x. Epub 2009 Dec 18. Clin Orthop Relat Res. 2010. PMID: 20020335 Free PMC article.
-
Outcome and serum ion determination up to 11 years after implantation of a cemented metal-on-metal hip prosthesis.Acta Orthop. 2009 Apr;80(2):168-73. doi: 10.3109/17453670902947408. Acta Orthop. 2009. PMID: 19404797 Free PMC article.
-
Long-Term Results of a Second-Generation, Small-Diameter, Metal-On-Metal Bearing in Primary Total Hip Arthroplasty at 14-year Follow-Up.Materials (Basel). 2020 Jan 24;13(3):557. doi: 10.3390/ma13030557. Materials (Basel). 2020. PMID: 31991553 Free PMC article.
-
Revision total hip arthroplasty for metal-on-metal failure.J Clin Orthop Trauma. 2020 Jan-Feb;11(1):9-15. doi: 10.1016/j.jcot.2019.09.021. Epub 2019 Oct 4. J Clin Orthop Trauma. 2020. PMID: 32001977 Free PMC article. Review.
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Medical