Polyethylene wear and acetabular component orientation
- PMID: 14652394
- DOI: 10.2106/00004623-200300004-00007
Polyethylene wear and acetabular component orientation
Abstract
Background: Polyethylene wear contributes substantially to both periprosthetic osteolysis and aseptic loosening after total hip arthroplasty. Acetabular component orientation has been shown to affect the range of motion of the hip as well as contact stresses. A series of studies was designed to test the hypothesis that acetabular component orientation can affect the magnitude and direction of polyethylene wear.
Methods: A finite-element model was used to compute contact stresses during a normal gait cycle. Wear at the end of each gait cycle was calculated with use of the sliding-distance-coupled finite-element formulation. The wear that was calculated with use of finite-element analysis was validated by comparison with the findings of hip wear simulator studies with the acetabular liner oriented to simulate 45 degrees and 55 degrees of abduction. In a clinical study, fifty-six patients who underwent sixty hip arthroplasties with use of a single prosthetic design were followed for as long as five years. Radiographs were analyzed to measure the abduction angle of the acetabular component and polyethylene wear.
Results: The finite-element analysis predicted increased peak contact stresses with an increased abduction angle and reduced peak contact stresses with an increased anteversion angle. Linear wear rates ranging from 0.036 to 0.045 mm/million cycles were also predicted, and increased acetabular abduction angles were predicted to be associated with higher linear wear rates. In the hip wear simulator studies, significantly different wear rates were found between the cups with acetabular abduction angles of 45 degrees and 55 degrees (mean, 17.2 compared with 21.7 mg/million cycles; p < 0.01). In the clinical study, radiographic analysis revealed significant correlation between the acetabular abduction angle and the linear polyethylene wear rate. A 40% increase in mean linear polyethylene wear was seen in cups with an abduction angle of >or=45 degrees. The direction of wear was more medial (by 9.4 degrees ) in cups with an abduction angle of <45 degrees.
Conclusion: All three studies presented here underlined the importance of optimizing the position of the acetabular component. Careful attention to acetabular position may help to minimize wear.
Similar articles
-
Effect of acetabular cup abduction angle on wear of ultrahigh-molecular-weight polyethylene in hip simulator testing.Am J Orthop (Belle Mead NJ). 2014 Oct;43(10):466-71. Am J Orthop (Belle Mead NJ). 2014. PMID: 25303445
-
Correlation of Cup Inclination Angle with Liner Wear for Metal-on-polyethylene in Hip Primary Arthroplasty.Orthop Surg. 2017 May;9(2):186-190. doi: 10.1111/os.12337. Epub 2017 May 30. Orthop Surg. 2017. PMID: 28557297 Free PMC article.
-
Highly crosslinked polyethylene wear rates and acetabular component orientation: a minimum ten-year follow-up.Bone Joint J. 2018 Jul;100-B(7):891-897. doi: 10.1302/0301-620X.100B7.BJJ-2017-1457.R3. Bone Joint J. 2018. PMID: 29954202
-
Dislocation after total hip arthroplasty: implant design and orientation.J Am Acad Orthop Surg. 2003 Mar-Apr;11(2):89-99. doi: 10.5435/00124635-200303000-00003. J Am Acad Orthop Surg. 2003. PMID: 12670135 Review.
-
The use of constrained components in total hip arthroplasty.J Am Acad Orthop Surg. 2002 Jul-Aug;10(4):233-8. doi: 10.5435/00124635-200207000-00001. J Am Acad Orthop Surg. 2002. PMID: 15089072 Review.
Cited by
-
[Prosthesis loosening].Z Rheumatol. 2012 Nov;71(9):785-97. doi: 10.1007/s00393-012-1042-4. Z Rheumatol. 2012. PMID: 23138556 German.
-
Upward and inward displacements of the acetabular component increase stress on femoral head in single endoprothesis models.Int Orthop. 2010 Apr;34(4):479-83. doi: 10.1007/s00264-009-0788-z. Epub 2009 May 8. Int Orthop. 2010. PMID: 19424694 Free PMC article.
-
Retrospective study comparing the accuracies of handheld infrared stereo camera and augmented reality-based navigation systems for total hip arthroplasty.Arch Orthop Trauma Surg. 2024 Jun;144(6):2811-2821. doi: 10.1007/s00402-024-05330-5. Epub 2024 May 5. Arch Orthop Trauma Surg. 2024. PMID: 38704436
-
Effects of body mass index and range of motion on intraoperative change in pelvic tilt during total hip arthroplasty using the direct anterior approach.BMC Musculoskelet Disord. 2021 Mar 2;22(1):240. doi: 10.1186/s12891-021-04087-x. BMC Musculoskelet Disord. 2021. PMID: 33653316 Free PMC article.
-
Central X-ray beam correction of radiographic acetabular cup measurement after THA: an experimental study.Int J Comput Assist Radiol Surg. 2017 May;12(5):829-837. doi: 10.1007/s11548-016-1489-x. Epub 2016 Oct 6. Int J Comput Assist Radiol Surg. 2017. PMID: 27714567
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Medical