Practice parameters for the use of portable monitoring devices in the investigation of suspected obstructive sleep apnea in adults
- PMID: 14655928
- DOI: 10.1093/sleep/26.7.907
Practice parameters for the use of portable monitoring devices in the investigation of suspected obstructive sleep apnea in adults
Abstract
Background: A variety of devices are used to evaluate patients with a potential diagnosis of obstructive sleep apnea (OSA). A committee comprised of members of the American Academy of Sleep Medicine, American Thoracic Society, and American College of Chest Physicians systematically evaluated data on the use of these devices and developed practice parameters.
Devices reviewed: Three categories of portable monitoring (PM) devices were reviewed with regard to assessing the probability of identifying an apnea-hypopnea index (AHI) of greater or less than 15 in attended and unattended settings. Type 2 (minimum of seven channels, including EEG, EOG, chin EMG, ECG or heart rate, airflow, respiratory effort, oxygen saturation), Type 3 (minimum of four channels, including ventilation or airflow (at least two channels of respiratory movement, or respiratory movement and airflow), heart rate or ECG and oxygen saturation) and Type 4 (most monitors of this type measure a single parameter or two parameters) devices were evaluated, and in-laboratory, attended polysomnography was used as a reference.
Specific recommendations: (1) Insufficient evidence is available to recommend the use of Type 2 PM devices in attended or unattended settings. (2) Type 3 PM devices appear to be capable of being used in an attended setting to increase or to decrease the probability that a patient has an apnea-hypopnea index greater than 15. (3) The use of Type 3 PM devices in an unattended setting is not recommended to rule in, rule out, or both rule in and rule out a diagnosis of OSA. (4) There is some evidence that the use of Type 3 PM devices in an attended in-laboratory setting may be acceptable to both rule in and rule out a diagnosis of OSA if certain limitations are in place. These limitations include manually scoring the records, using the devices only in patients without significant comorbid conditions, having an awareness that symptomatic patients with a negative study should have a Type 1 study, and not using these devices for titrating positive airway pressure or conducting split-night studies. (5) The use of Type 4 PM devices in attended or unattended settings is not recommended.
General recommendations: Type 3 and 4 PM devices cannot score sleep and, therefore, do not meet some current Medicare guidelines. The use of PM devices is not recommended for general-population screening or in the absence of a pretest probability of the patient having a diagnosis of OSA, for complaints other than those associated with OSA, without review of raw data during interpretation, by physicians without familiarity with their use and limitations, and without trained personnel to perform technical scoring. Future research should address the use of PM devices in patients with comorbid conditions; non-White patients and women; larger, better-controlled studies; studies focused on the use of Type 2 and 3 devices; studies focusing on decision making and outcomes rather than simple classification using arbitrary cutoffs; and studies that seek to elucidate cost-effectiveness data on the use of PM devices.
Similar articles
-
Effectiveness of Portable Monitoring Devices for Diagnosing Obstructive Sleep Apnea: Update of a Systematic Review [Internet].Rockville (MD): Agency for Healthcare Research and Quality (US); 2004 Sep 1. Rockville (MD): Agency for Healthcare Research and Quality (US); 2004 Sep 1. PMID: 26065047 Free Books & Documents. Review.
-
Clinical guidelines for the use of unattended portable monitors in the diagnosis of obstructive sleep apnea in adult patients. Portable Monitoring Task Force of the American Academy of Sleep Medicine.J Clin Sleep Med. 2007 Dec 15;3(7):737-47. J Clin Sleep Med. 2007. PMID: 18198809 Free PMC article.
-
Practice parameters for the use of autotitrating continuous positive airway pressure devices for titrating pressures and treating adult patients with obstructive sleep apnea syndrome: an update for 2007. An American Academy of Sleep Medicine report.Sleep. 2008 Jan;31(1):141-7. doi: 10.1093/sleep/31.1.141. Sleep. 2008. PMID: 18220088 Free PMC article.
-
Practice parameters for the use of auto-titrating continuous positive airway pressure devices for titrating pressures and treating adult patients with obstructive sleep apnea syndrome. An American Academy of Sleep Medicine report.Sleep. 2002 Mar 15;25(2):143-7. doi: 10.1093/sleep/25.2.143. Sleep. 2002. PMID: 11902424
-
Utility of portable monitoring in the diagnosis of obstructive sleep apnea.J Postgrad Med. 2015 Oct-Dec;61(4):223-9. doi: 10.4103/0022-3859.166509. J Postgrad Med. 2015. PMID: 26440391 Free PMC article. Review.
Cited by
-
Experimental pain and opioid analgesia in volunteers at high risk for obstructive sleep apnea.PLoS One. 2013;8(1):e54807. doi: 10.1371/journal.pone.0054807. Epub 2013 Jan 29. PLoS One. 2013. PMID: 23382975 Free PMC article. Clinical Trial.
-
Continuous positive airway pressure device-based automated detection of obstructive sleep apnea compared to standard laboratory polysomnography.Sleep Breath. 2010 Jun;14(2):101-7. doi: 10.1007/s11325-009-0285-z. Epub 2009 Oct 14. Sleep Breath. 2010. PMID: 19826848
-
[Detecting of obstructive sleep apnea hypopnea syndrome using a multi-parameter pressure sensitive sleep monitor].Beijing Da Xue Xue Bao Yi Xue Ban. 2021 Jun 18;53(3):608-612. doi: 10.19723/j.issn.1671-167X.2021.03.029. Beijing Da Xue Xue Bao Yi Xue Ban. 2021. PMID: 34145870 Free PMC article. Chinese.
-
Obstructive sleep apnea as a potential risk factor for aortic disease.Heart Vessels. 2012 Mar;27(2):166-73. doi: 10.1007/s00380-011-0135-3. Epub 2011 Mar 26. Heart Vessels. 2012. PMID: 21442254
-
Entering the age of portable monitoring.J Clin Sleep Med. 2008 Aug 15;4(4):303-4. J Clin Sleep Med. 2008. PMID: 18763419 Free PMC article. No abstract available.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources