The accuracy of portable peak flow meters
- PMID: 1465746
- PMCID: PMC464089
- DOI: 10.1136/thx.47.11.904
The accuracy of portable peak flow meters
Abstract
Background: The variability of peak expiratory flow (PEF) is now commonly used in the diagnosis and management of asthma. It is essential for PEF meters to have a linear response in order to obtain an unbiased measurement of PEF variability. As the accuracy and linearity of portable PEF meters have not been rigorously tested in recent years this aspect of their performance has been investigated.
Methods: The response of several portable PEF meters was tested with absolute standards of flow generated by a computer driven, servo controlled pump and their response was compared with that of a pneumotachograph.
Results: For each device tested the readings were highly repeatable to within the limits of accuracy with which the pointer position can be assessed by eye. The between instrument variation in reading for six identical devices expressed as a 95% confidence limit was, on average across the range of flows, +/- 8.5 l/min for the Mini-Wright, +/- 7.9 l/min for the Vitalograph, and +/- 6.4 l/min for the Ferraris. PEF meters based on the Wright meter all had similar error profiles with overreading of up to 80 l/min in the mid flow range from 300 to 500 l/min. This overreading was greatest for the Mini-Wright and Ferraris devices, and less so for the original Wright and Vitalograph meters. A Micro-Medical Turbine meter was accurate up to 400 l/min and then began to underread by up to 60 l/min at 720 l/min. For the low range devices the Vitalograph device was accurate to within 10 l/min up to 200 l/min, with the Mini-Wright overreading by up to 30 l/min above 150 l/min.
Conclusion: Although the Mini-Wright, Ferraris, and Vitalograph meters gave remarkably repeatable results their error profiles for the full range meters will lead to important errors in recording PEF variability. This may lead to incorrect diagnosis and bias in implementing strategies of asthma treatment based on PEF measurement.
Similar articles
-
Inadequate peak expiratory flow meter characteristics detected by a computerised explosive decompression device.Thorax. 2003 May;58(5):411-6. doi: 10.1136/thorax.58.5.411. Thorax. 2003. PMID: 12728161 Free PMC article.
-
The Peak Flow Working Group: test of portable peak flow meters by explosive decompression.Eur Respir J Suppl. 1997 Feb;24:23S-25S. Eur Respir J Suppl. 1997. PMID: 9098705
-
Frequency response of variable orifice type peak flow meters: requirements and testing.Eur Respir J. 1995 May;8(5):849-55. Eur Respir J. 1995. PMID: 7656961
-
Incorrect use of peak flow meters: are you observing your patients?J Asthma. 2014 Aug;51(6):566-72. doi: 10.3109/02770903.2014.914218. Epub 2014 May 9. J Asthma. 2014. PMID: 24720711 Review.
-
[Comparison of different types of peak flow meter and reference values for peak expiratory flow (PEF) for healthy Japanese subjects].Nihon Rinsho. 1996 Nov;54(11):2927-32. Nihon Rinsho. 1996. PMID: 8950932 Review. Japanese.
Cited by
-
Effect of altitude on spirometric parameters and the performance of peak flow meters.Thorax. 1996 Feb;51(2):175-8. doi: 10.1136/thx.51.2.175. Thorax. 1996. PMID: 8711651 Free PMC article.
-
Potential effects of correction of inaccuracies of the mini-Wright peak expiratory flow meter on the use of an asthma self-management plan.Thorax. 1996 Apr;51(4):403-6. doi: 10.1136/thx.51.4.403. Thorax. 1996. PMID: 8733493 Free PMC article.
-
Peak flow measurement.Thorax. 1992 Nov;47(11):903. doi: 10.1136/thx.47.11.903. Thorax. 1992. PMID: 1465745 Free PMC article. No abstract available.
-
Peak expiratory flow rate underestimates severity of airflow obstruction in acute asthma.Korean J Intern Med. 2002 Sep;17(3):174-9. doi: 10.3904/kjim.2002.17.3.174. Korean J Intern Med. 2002. PMID: 12298428 Free PMC article.
-
Development of an expert system for the interpretation of serial peak expiratory flow measurements in the diagnosis of occupational asthma. Midlands Thoracic Society Research Group.Occup Environ Med. 1999 Nov;56(11):758-64. doi: 10.1136/oem.56.11.758. Occup Environ Med. 1999. PMID: 10658562 Free PMC article.
References
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Medical
Research Materials