Skip to main page content
U.S. flag

An official website of the United States government

Dot gov

The .gov means it’s official.
Federal government websites often end in .gov or .mil. Before sharing sensitive information, make sure you’re on a federal government site.

Https

The site is secure.
The https:// ensures that you are connecting to the official website and that any information you provide is encrypted and transmitted securely.

Access keys NCBI Homepage MyNCBI Homepage Main Content Main Navigation
. 2003 Sep;13(3):358-68.
doi: 10.1123/ijsnem.13.3.358.

Accuracy of three dry-chemistry methods for lipid profiling and risk factor classification

Affiliations

Accuracy of three dry-chemistry methods for lipid profiling and risk factor classification

Daniela A Rubin et al. Int J Sport Nutr Exerc Metab. 2003 Sep.

Abstract

The purpose of this project was to determine the accuracy in lipids measurement and risk factor classification using Reflotron, Cholestech, and Ektachem DT-60 dry-chemistry analyzers. Plasma and capillary venous blood from fasting subjects (n = 47) were analyzed for total cholesterol (TC), high density lipoprotein (HDL-C), and triglycerides (TG) using these analyzers and a CDC certified laboratory. Accuracy was evaluated by comparing the results of each portable analyzer against the CDC reference method. One-way ANOVAs were performed for TC, HDL-C, and TG between all portable analyzers and the reference method. Chi-square was used for risk classification (2001 NIH Guidelines). Compared to the reference method, the Ektachem and Reflotron provided significantly lower values for TC (p < .05). In addition, the Cholestech and Ektachem values for HDL-C were higher than the CDC (p < .05). The Reflotron and Cholestech provided higher values of TG than the CDC (p < .05). Chi-squares analyses for risk classification were not significant (p > .45) between analyzers. According to these results, the Ektachem and Cholestech analyzers met the current NCEP III guidelines for accuracy in measurement of TC, while only Ektachem met guidelines for TG. All 3 analyzers provided a good overall risk classification; however, values of HDL-C should be only used for screening purposes.

PubMed Disclaimer

Publication types

LinkOut - more resources