Health care workers: risk factors for nonlatex and latex gloves during surgery
- PMID: 14674793
- DOI: 10.1202/542.1
Health care workers: risk factors for nonlatex and latex gloves during surgery
Abstract
This study examined glove failure and related factors in both nonlatex and latex surgical gloves after routine use. A federally funded research study was conducted to collect surgical gloves from those directly involved in surgical procedures. All gloves were examined in the laboratory for both visual defects and barrier integrity. A total of 11,118 usable surgical gloves were examined. The overall defect rate was 7.8%; nonlatex gloves were significantly more likely to fail (8.4%) than latex gloves (6.9%). The majority of defects in the latex gloves (90%) and nonlatex gloves (70%) were not detected by visual examination. Separate logistic regression models examined predictors of defects for the gloves. The only factor that increased the odds of a defect for a latex glove was duration of use over 6 hours. Factors increasing the odds of a defect in nonlatex gloves included gloves worn by a scrub person and gloves used in certain surgical services. Scrub persons had a higher defect rate despite wearing their gloves for a significantly shorter time than other health care workers. Latex and nonlatex gloves fail under different conditions. Latex gloves fail primarily due to length of use, whereas nonlatex gloves are more sensitive to conditions of us (e.g., type of health care worker and type of surgery). Providers can help guard against glove defects by double gloving and by changing gloves often, especially when using nonlatex gloves in higher-risk surgeries.
Similar articles
-
Failure rates in nonlatex surgical gloves.Am J Infect Control. 2004 Aug;32(5):268-73. doi: 10.1016/j.ajic.2003.12.005. Am J Infect Control. 2004. PMID: 15292890
-
[What is the appropriate time-interval for changing gloves during surgical procedures].Zentralbl Chir. 2010 Feb;135(1):25-7. doi: 10.1055/s-0029-1224684. Epub 2010 Jan 27. Zentralbl Chir. 2010. PMID: 20108181 Review. German.
-
A laboratory-based study to assess the performance of surgical gloves.AORN J. 2003 Apr;77(4):772-9. doi: 10.1016/s0001-2092(06)60796-5. AORN J. 2003. PMID: 12705733
-
[Defects in the surgical glove barrier. Single or double gloves].Ugeskr Laeger. 2003 Mar 3;165(10):1016-9. Ugeskr Laeger. 2003. PMID: 12645406 Review. Danish.
-
Glove safety: summary of recent findings and recommendations from health care regulators.South Med J. 1995 Nov;88(11):1093-8. South Med J. 1995. PMID: 7481977 Review.