Measuring eye aberrations with Hartmann-Shack wave-front sensors: should the irradiance distribution across the eye pupil be taken into account?
- PMID: 14686502
- DOI: 10.1364/josaa.20.002237
Measuring eye aberrations with Hartmann-Shack wave-front sensors: should the irradiance distribution across the eye pupil be taken into account?
Abstract
A usual approximation in Hartmann-Shack aberrometry is that the centroid displacements are proportional to the spatial averages of the wave-front slopes at the sampling subapertures. However, these spatial averages are actually weighted by the local irradiance distribution across each microlens. The irradiance across the eye pupil is not uniform in usual reflectometric aberrometers, which is due to several factors including retinal scattering and cone waveguiding directionality. It is shown that neglecting this fact in usual least-squares reconstruction procedures gives rise to a biased estimation of the aberration coefficients. The magnitude of this bias depends on the actual irradiance distribution across the eye pupil, the mode being estimated, the detailed modal composition of the aberrated wave front, and the geometry of the wave-front sampling array. Order-of-magnitude calculations suggest that this bias may well be in the range 5%-10% for relatively smooth irradiance distributions. The systematic nature of this error makes it advisable to check for its presence and, if required, to compensate for it by an adequate choice of the least-squares reconstruction matrix.
Similar articles
-
Ocular aberrations with ray tracing and Shack-Hartmann wave-front sensors: does polarization play a role?J Opt Soc Am A Opt Image Sci Vis. 2002 Jun;19(6):1063-72. doi: 10.1364/josaa.19.001063. J Opt Soc Am A Opt Image Sci Vis. 2002. PMID: 12049343
-
Comparison of the eye's wave-front aberration measured psychophysically and with the Shack-Hartmann wave-front sensor.J Opt Soc Am A Opt Image Sci Vis. 1998 Sep;15(9):2457-65. doi: 10.1364/josaa.15.002457. J Opt Soc Am A Opt Image Sci Vis. 1998. PMID: 9729857
-
Measuring ocular aberrations and image quality in peripheral vision with a clinical wavefront aberrometer.Clin Exp Optom. 2009 May;92(3):212-22. doi: 10.1111/j.1444-0938.2009.00376.x. Clin Exp Optom. 2009. PMID: 19462503
-
Aberrometry: basic science and clinical applications.Bull Soc Belge Ophtalmol. 2006;(302):197-213. Bull Soc Belge Ophtalmol. 2006. PMID: 17265799 Review.
-
Principles of Hartmann-Shack aberrometry.J Refract Surg. 2000 Sep-Oct;16(5):S563-5. doi: 10.3928/1081-597X-20000901-14. J Refract Surg. 2000. PMID: 11019873 Review. No abstract available.
Cited by
-
Accounting for focal shift in the Shack-Hartmann wavefront sensor.Opt Lett. 2019 Sep 1;44(17):4151-4154. doi: 10.1364/OL.44.004151. Opt Lett. 2019. PMID: 31465350 Free PMC article.
-
Centroid error due to non-uniform lenslet illumination in the Shack-Hartmann wavefront sensor.Opt Lett. 2019 Sep 1;44(17):4167-4170. doi: 10.1364/OL.44.004167. Opt Lett. 2019. PMID: 31465354 Free PMC article.
-
Wavefront sensing and reconstruction from gradient and Laplacian data measured with a Hartmann-Shack sensor.Opt Lett. 2006 Jun 15;31(12):1845-7. doi: 10.1364/ol.31.001845. Opt Lett. 2006. PMID: 16729090 Free PMC article.
-
Average gradient of Zernike polynomials over polygons.Opt Express. 2020 Jun 22;28(13):18876-18886. doi: 10.1364/OE.393223. Opt Express. 2020. PMID: 32672177 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Medical