Eccentricity and measurement variability and repeatability with the retinal thickness analyser
- PMID: 14693775
- PMCID: PMC1771935
- DOI: 10.1136/bjo.88.1.62
Eccentricity and measurement variability and repeatability with the retinal thickness analyser
Abstract
Aims: To define the variability and repeatability of retinal thickness measurements using the retinal thickness analyser (RTA) and to elucidate any interaction between eccentricity (that is, position relative to the fovea) and variability and repeatability.
Methods: The sample comprised 20 normal subjects of mean age 33 years. Each subject attended for two visits. Repeated RTA scans were acquired centred on the fovea and for any one of the four possible non-foveal scan areas. The mean retinal thickness (+SD) was calculated for a series of concentric circular bands centred on fixation. A repeated measures analysis of variance (ANOVA) was used to determine any significant interaction between the variability of RTA thickness values and eccentricity.
Results: The group mean coefficient of variation and coefficient of repeatability were highest at the fovea. The repeated measures ANOVA revealed that the within test variability of RTA measurements varied significantly with eccentricity (p<0.0001). Similarly, the between test repeatability varied significantly with eccentricity (p = 0.045).
Conclusion: The significantly elevated within test variability and between test repeatability in the foveal area need to be considered when using the RTA to evaluate patients with macular disease.
Figures


Similar articles
-
Repeatability of Foveal Measurements Using Spectralis Optical Coherence Tomography Segmentation Software.PLoS One. 2015 Jun 15;10(6):e0129005. doi: 10.1371/journal.pone.0129005. eCollection 2015. PLoS One. 2015. PMID: 26076457 Free PMC article.
-
Retinal thickness measurements with optical coherence tomography and the scanning retinal thickness analyzer.Retina. 2001;21(1):57-61. doi: 10.1097/00006982-200102000-00010. Retina. 2001. PMID: 11217931
-
Comparison of foveal thickness measured with the retinal thickness analyzer and optical coherence tomography.Retina. 2001;21(6):596-601. doi: 10.1097/00006982-200112000-00006. Retina. 2001. PMID: 11756882
-
Retinal thickness variation in the diabetic patient measured by the retinal thickness analyser.Br J Ophthalmol. 1998 Sep;82(9):1003-6. doi: 10.1136/bjo.82.9.1003. Br J Ophthalmol. 1998. PMID: 9893588 Free PMC article.
-
[New examination methods for macular disorders--application of diagnosis and treatment].Nippon Ganka Gakkai Zasshi. 2000 Dec;104(12):899-942. Nippon Ganka Gakkai Zasshi. 2000. PMID: 11193944 Review. Japanese.
Cited by
-
Repeatability of Foveal Measurements Using Spectralis Optical Coherence Tomography Segmentation Software.PLoS One. 2015 Jun 15;10(6):e0129005. doi: 10.1371/journal.pone.0129005. eCollection 2015. PLoS One. 2015. PMID: 26076457 Free PMC article.
-
A reference standard for the measurement of macular oedema.Br J Ophthalmol. 2006 Sep;90(9):1197-202. doi: 10.1136/bjo.2006.095885. Br J Ophthalmol. 2006. PMID: 16929064 Free PMC article. Review.
-
Optical coherence tomography (OCT) in neuro-ophthalmology.Eye (Lond). 2021 Jan;35(1):17-32. doi: 10.1038/s41433-020-01288-x. Epub 2020 Nov 25. Eye (Lond). 2021. PMID: 33239763 Free PMC article. Review.
References
-
- Hudson C, Flanagan JG, McLeod D. A clinical vision science perspective of the management of diabetic macular oedema. Focus on diabetic retinopathy. Ex Med 1999;6:41–4.
-
- Shahidi M, Zeimer R, Mori M. Topography of the retinal thickness in normal subjects. Ophthalmology 1990;97:1120–4. - PubMed
-
- Zeimer R, Shahidi M, Mori M, et al. A new method for rapid mapping of the retinal thickness at the posterior pole. Invest Ophthalmol Vis Sci 1996;37:1994–2001. - PubMed
-
- Landau D, Schneidman EM, Jacobovitz T, et al.. Quantitative in vivo retinal thickness measurements in healthy subjects. Ophthalmology 1997;104:639–42. - PubMed
-
- Asrani S, Zou S, d’Anna S, et al. Noninvasive mapping of the normal retinal thickness at the posterior pole. Ophthalmology 1999;106:269–73. - PubMed
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources