A new antiglycolytic agent
- PMID: 14713384
- DOI: 10.1258/000456304322664681
A new antiglycolytic agent
Abstract
Background: Glycolysis is not completely or predictably inhibited by the glucose preservative currently in use, with glucose values falling by as much as 0.5 mmol/L during a 2-4-h period after sample collection. Immediate centrifugation of all samples is also impractical and therefore misdiagnosis of disease can occur, especially if more emphasis is being placed on fasting glucose for the diagnosis of diabetes.
Methods: Glycolysis at room temperature was evaluated over time using glyceraldehyde alone as well as in conjunction with standard antiglycolytic agents.
Results: Glyceraldehyde alone does not inhibit glycolysis completely. The combination of 11 mmol/L glyceraldehyde, 119 mmol/L sodium fluoride and 21.7 mmol/L potassium oxalate gave the best antiglycolytic results. The glucose values measured in samples stored at room temperature for 48 h was no different from those measured in samples centrifuged immediately after venepuncture and this is clinically superior to conventionally used sodium fluoride and potassium oxalate.
Conclusion: Plasma glucose concentrations obtained from blood collected into tubes containing glyceraldehyde, sodium fluoride and potassium oxalate will more closely reflect those of the patient at venepuncture.
Similar articles
-
Rapid blood separation is superior to fluoride for preventing in vitro reductions in measured blood glucose concentration.J Clin Pathol. 2009 Aug;62(8):752-3. doi: 10.1136/jcp.2008.062547. J Clin Pathol. 2009. PMID: 19638548
-
[Erroneous diagnosis 'neonatal hypoglycemia' due to incorrect preservation of blood samples].Ned Tijdschr Geneeskd. 1991 Sep 14;135(37):1691-4. Ned Tijdschr Geneeskd. 1991. PMID: 1922513 Dutch.
-
Effects of standard anticoagulants and storage procedures on plasma glucose values in seals.J Am Vet Med Assoc. 1992 Jul 1;201(1):145-8. J Am Vet Med Assoc. 1992. PMID: 1644639
-
Blood Glucose Determination: Effect of Tube Additives.Adv Clin Chem. 2018;84:101-123. doi: 10.1016/bs.acc.2017.12.003. Epub 2018 Feb 1. Adv Clin Chem. 2018. PMID: 29478513
-
Importance of the pre-analytical phase in blood glucose analysis.Acta Clin Belg. 2010 Sep-Oct;65(5):311-8. doi: 10.1179/acb.2010.068. Acta Clin Belg. 2010. PMID: 21128557 Review.
Cited by
-
Optimizing d-mannose and glyceraldehyde concentrations as glucose preservatives without clinically affecting biochemical test results.Pract Lab Med. 2024 Mar 13;39:e00388. doi: 10.1016/j.plabm.2024.e00388. eCollection 2024 Mar. Pract Lab Med. 2024. PMID: 38623088 Free PMC article.
-
Random blood glucose measurement at antenatal booking to screen for overt diabetes in pregnancy: a retrospective study.Diabetes Care. 2011 Oct;34(10):2217-9. doi: 10.2337/dc11-0533. Epub 2011 Aug 15. Diabetes Care. 2011. PMID: 21844290 Free PMC article.
-
Antitumor Effect of Morusin via G1 Arrest and Antiglycolysis by AMPK Activation in Hepatocellular Cancer.Int J Mol Sci. 2021 Sep 30;22(19):10619. doi: 10.3390/ijms221910619. Int J Mol Sci. 2021. PMID: 34638959 Free PMC article.
-
Accuracy of the StatStrip versus SureStep Flexx glucose meter in neonates at risk of hypoglycemia.Eur J Pediatr. 2013 Sep;172(9):1181-6. doi: 10.1007/s00431-013-2019-2. Epub 2013 May 5. Eur J Pediatr. 2013. PMID: 23644649 Clinical Trial.
-
Glycolysis inhibitors negatively bias blood glucose measurements: potential impact on the reported prevalence of diabetes mellitus.J Clin Pathol. 2007 Jul;60(7):820-3. doi: 10.1136/jcp.2006.039925. J Clin Pathol. 2007. PMID: 17596547 Free PMC article.
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical