Multicenter pivotal trial results of the Lifepath System for endovascular aortic aneurysm repair
- PMID: 14718809
- DOI: 10.1016/j.jvs.2003.10.036
Multicenter pivotal trial results of the Lifepath System for endovascular aortic aneurysm repair
Abstract
Purpose: This study was undertaken to assess the results of endovascular aortic aneurysm repair with the Lifepath abdominal aortic aneurysm (AAA) graft system.
Method: In a prospective clinical trial, 23 centers used the Lifepath System balloon-expandable, modular bifurcated stent graft for elective endovascular aortic aneurysm repair. Stent grafts were sized according to computed tomographic angiography-based diameter measurements. All repairs were performed in the operating room through bilateral surgically exposed femoral arteries. Results were assessed with contrast agent-enhanced computed tomography scans and plain abdominal x-ray films at 1, 6, 12, 24, 36, and 48 months postoperatively.
Results: Over 52 months (mean follow-up, 11 months), 227 patients (206 men, 21 women) were enrolled. Technical implant success rate was 98.7%. There were five (2.2%) conversions to open surgery: two emergently because of aortic perforation; to treat refractory endoleak, immediate in one and at 12 months in one; and to replace a device with wireform fractures that had migrated at 12 months, resulting in a proximal endoleak. The perioperative mortality rate was 1.3%. There was one operative death during a secondary procedure to repair perforation of the aorta. There were two perioperative deaths, from postoperative myocardial infarction (n = 1) and pulmonary embolus (n = 1). There were 12 late deaths, from coronary artery disease (n = 4), cancer (n = 2), respiratory failure (n = 2), sepsis (n = 1), or unknown cause (n = 3). Median length of stay was 2 days (mean, 4 days). There have been no AAA ruptures after successful implantation of the device, no graft limb thromboses, and no limb dislocations. At the time of operation endoleak was noted in 43 (19%) patients, but by 6 months this was reduced to 8 (5.9%) patients (type I, n = 1; type II, n = 7). There were no type III or type IV endoleaks. Secondary interventions to treat endoleaks included open conversion (n = 2), placement of extension cuffs (n = 4), repeated balloon dilation (n = 3), and coil embolization (n = 6). The two remaining secondary interventions were emergent treatment of postoperative bleeding from a groin incision, and a colon resection because of postoperative colonic ischemia, for a 12-month secondary intervention rate of 7.5%. Wireform fractures were noted in the first generation Lifepath device in 37 of 79 (47%) patients. Graft migration (>10 mm) was observed in five patients (2.2%), each of whom also had two or more fractures of the proximal anchoring wireforms. Migration resulted in a proximal attachment endoleak in one patient. In response to wireform fractures, the device was modified after the initial 79 patients were enrolled. Wireform fracture has been observed in six patients since this modification (4%), and in only one patient did this involve fracture of a proximal anchoring wireform; none of these patients has had endoleak or graft migration. By 12 months, mean aneurysm diameter was noted to decrease by 9 mm (P <.0001), and mean aneurysm volume by 42 mL (P <.0001) from the preoperative visit.
Conclusion: The Lifepath System demonstrates a low endoleak and secondary intervention rate and high sac regression rate, compared with other devices. The unique balloon-expandable design offers the advantages of precise placement and high radial force. The device appears to be highly resistant to limb thrombosis and modular component separation. Patients were protected from AAA rupture after successful device implantation, and demonstrated significant reduction in AAA diameter and volume. Fractures of the wireforms of the main body of the device have been observed. Careful long-term follow-up is necessary.
Similar articles
-
Midterm results of the multicenter trial of the powerlink bifurcated system for endovascular aortic aneurysm repair.J Vasc Surg. 2004 Nov;40(5):849-59. doi: 10.1016/j.jvs.2004.08.051. J Vasc Surg. 2004. PMID: 15557896 Clinical Trial.
-
Outcomes of secondary interventions after abdominal aortic aneurysm endovascular repair.J Vasc Surg. 2004 Feb;39(2):298-305. doi: 10.1016/j.jvs.2003.09.043. J Vasc Surg. 2004. PMID: 14743128
-
[Endovascular therapy of abdominal aortic aneurysm: results of a mid-term follow-up].Rofo. 2003 Oct;175(10):1392-402. doi: 10.1055/s-2003-42881. Rofo. 2003. PMID: 14556109 German.
-
The chimney graft technique for preserving visceral vessels during endovascular treatment of aortic pathologies.J Vasc Surg. 2012 May;55(5):1497-503. doi: 10.1016/j.jvs.2011.10.009. Epub 2012 Jan 10. J Vasc Surg. 2012. PMID: 22236883 Review.
-
Flow-diverting stents for the treatment of arterial aneurysms.J Vasc Surg. 2012 Sep;56(3):839-46. doi: 10.1016/j.jvs.2012.04.020. Epub 2012 Jul 26. J Vasc Surg. 2012. PMID: 22840737 Review.
Cited by
-
Acute ischemia secondary to popliteal artery stent fracture and embolization.J Vasc Surg Cases Innov Tech. 2023 Apr 23;9(3):101143. doi: 10.1016/j.jvscit.2023.101143. eCollection 2023 Sep. J Vasc Surg Cases Innov Tech. 2023. PMID: 37799850 Free PMC article.
-
Surveillance Imaging Following Endovascular Aneurysm Repair.Semin Intervent Radiol. 2015 Sep;32(3):239-48. doi: 10.1055/s-0035-1556878. Semin Intervent Radiol. 2015. PMID: 26327742 Free PMC article. Review.
-
Limb graft occlusion following endovascular aortic repair: Incidence, causes, treatment and prevention in a study cohort.Exp Ther Med. 2017 Aug;14(2):1763-1768. doi: 10.3892/etm.2017.4658. Epub 2017 Jun 23. Exp Ther Med. 2017. PMID: 28810647 Free PMC article.
-
Surveillance Imaging following Endovascular Aneurysm Repair: State of the Art.Semin Intervent Radiol. 2020 Oct;37(4):356-364. doi: 10.1055/s-0040-1715882. Epub 2020 Oct 1. Semin Intervent Radiol. 2020. PMID: 33041481 Free PMC article. Review.
-
Prevalence and risk factors of type II endoleaks after endovascular aneurysm repair: A meta-analysis.PLoS One. 2017 Feb 9;12(2):e0170600. doi: 10.1371/journal.pone.0170600. eCollection 2017. PLoS One. 2017. PMID: 28182753 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources