Entropy of EEG during anaesthetic induction: a comparative study with propofol or nitrous oxide as sole agent
- PMID: 14722164
- DOI: 10.1093/bja/aeh036
Entropy of EEG during anaesthetic induction: a comparative study with propofol or nitrous oxide as sole agent
Abstract
Background: The search continues for an anaesthetic monitor that can define the level of anaesthesia in an individual patient irrespective of anaesthetic agent(s) used. Studies of available monitors based on bispectral analysis or evoked auditory potentials show the complexity of the problem. We assessed a new monitor, based on the entropy of the EEG, during induction of anaesthesia with either propofol or nitrous oxide.
Methods: In an open, randomized study (two groups; n=10) of day surgical patients, we induced loss of response with incremental boluses of propofol. The other group was given propofol 30 mg and then increasing concentrations of nitrous oxide until loss of response. We measured entropy with the M-Entropy Module S/5 (Datex-Ohmeda) using forehead electrodes and recorded response entropy (RE; including frontal electromyogram) and state entropy (SE; only the cortical EEG). Values are median (range).
Results: Baseline values were RE 98 (96-100), SE 89 (87-91) and RE 98 (96-99), SE 89 (87-91) for the propofol and nitrous oxide patients, respectively. During propofol induction, both entropy indices decreased with increasing sedation, with RE 40 (23-76) and SE 34 (17-70) at loss of response. Neither RE nor SE decreased during nitrous oxide inhalation, and at loss of response using nitrous oxide, RE and SE were unchanged at 98 (96-100) and 88 (85-91) respectively.
Conclusions: The entropy monitor of anaesthetic depth shows a successive decrease with propofol but loss of consciousness with nitrous oxide is not associated with change in entropy indices.
Comment in
-
Entropy is blind to nitrous oxide. Can we see why?Br J Anaesth. 2004 Feb;92(2):159-61. doi: 10.1093/bja/aeh039. Br J Anaesth. 2004. PMID: 14722161 No abstract available.
Similar articles
-
Facial muscle activity, Response Entropy, and State Entropy indices during noxious stimuli in propofol-nitrous oxide or propofol-nitrous oxide-remifentanil anaesthesia without neuromuscular block.Br J Anaesth. 2009 Feb;102(2):227-33. doi: 10.1093/bja/aen356. Epub 2008 Dec 25. Br J Anaesth. 2009. PMID: 19112059 Free PMC article.
-
Cerebral state index during anaesthetic induction: a comparative study with propofol or nitrous oxide.Acta Anaesthesiol Scand. 2005 Jul;49(6):750-3. doi: 10.1111/j.1399-6576.2005.00737.x. Acta Anaesthesiol Scand. 2005. PMID: 15954953 Clinical Trial.
-
Comparative evaluation of the Datex-Ohmeda S/5 Entropy Module and the Bispectral Index monitor during propofol-remifentanil anesthesia.Anesthesiology. 2004 Dec;101(6):1283-90. doi: 10.1097/00000542-200412000-00007. Anesthesiology. 2004. PMID: 15564934
-
Special cases: ketamine, nitrous oxide and xenon.Best Pract Res Clin Anaesthesiol. 2006 Mar;20(1):69-79. doi: 10.1016/j.bpa.2005.08.014. Best Pract Res Clin Anaesthesiol. 2006. PMID: 16634415 Review.
-
Entropy.Best Pract Res Clin Anaesthesiol. 2006 Mar;20(1):101-9. doi: 10.1016/j.bpa.2005.07.009. Best Pract Res Clin Anaesthesiol. 2006. PMID: 16634417 Review.
Cited by
-
High Variability Periods in the EEG Distinguish Cognitive Brain States.Brain Sci. 2023 Oct 30;13(11):1528. doi: 10.3390/brainsci13111528. Brain Sci. 2023. PMID: 38002488 Free PMC article.
-
General anesthetics and molecular mechanisms of unconsciousness.Int Anesthesiol Clin. 2008 Summer;46(3):43-53. doi: 10.1097/AIA.0b013e3181755da5. Int Anesthesiol Clin. 2008. PMID: 18617817 Free PMC article. Review.
-
Overview and Clinical Applications of Artificial Intelligence and Machine Learning in Cardiac Anesthesiology.J Cardiothorac Vasc Anesth. 2024 May;38(5):1211-1220. doi: 10.1053/j.jvca.2024.02.004. Epub 2024 Feb 15. J Cardiothorac Vasc Anesth. 2024. PMID: 38453558 Free PMC article. Review.
-
The effect of addition of nitrous oxide to a sevoflurane anesthetic on BIS, PSI, and entropy.J Clin Monit Comput. 2006 Jun;20(3):145-50. doi: 10.1007/s10877-006-9009-0. Epub 2006 Jun 21. J Clin Monit Comput. 2006. PMID: 16794892 Clinical Trial.
-
Population based models of cortical drug response: insights from anaesthesia.Cogn Neurodyn. 2008 Dec;2(4):283-96. doi: 10.1007/s11571-008-9063-z. Epub 2008 Sep 23. Cogn Neurodyn. 2008. PMID: 19003456 Free PMC article.
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Medical