Continuous intra-arterial oximetry, pulse oximetry, and co-oximetry during cardiac surgery
- PMID: 1472661
- DOI: 10.1016/1053-0770(92)90048-c
Continuous intra-arterial oximetry, pulse oximetry, and co-oximetry during cardiac surgery
Abstract
This study evaluated arterial catheter oximetry versus pulse oximetry in eight patients (ASA III-IV) who underwent cardiac surgery. Co-oximeter saturation values served as the standard. Arterial oxygen saturation was determined simultaneously with these three methods at 162 prospectively defined points of measurement before, during, and after cardiopulmonary bypass (CPB). At the same times before and after CPB, arterial, pulmonary arterial, and central venous pressures, and cardiac output determinations were recorded. Saturation readings were obtained in more than 99% of measurements with catheter oximetry and in only 59% to 84% of measurements with pulse oximetry. Failure of pulse oximetry correlated with low mean arterial pressures and low cardiac outputs, but not with high systemic vascular resistance. The mean saturation values determined by catheter oximetry as well as by pulse oximetry differed from the mean values obtained by co-oximetry by less than 1% (= bias). The standard deviations of the individual differences between readings of catheter or pulse oximetry and readings of co-oximetry (= precision) were +/- 0.5% to +/- 1.0% for catheter oximetry and +/- 1.0% to +/- 1.2% for pulse oximetry. In summary, catheter oximetry was superior to pulse oximetry with regard to obtaining readings and to reliability of the obtained readings. Invasiveness and high costs influence the decision as to whether to use catheter oximetry, but if reliable and precise measurements of saturation are important at any time during surgery, pulse oximetry is an insufficient method and co-oximetry is a time-consuming method of analysis, whereas catheter oximetry is quick, reliable, and precise.
Similar articles
-
[Intra-arterial catheter oximetry and pulse oximetry in comparison with CO-oximetry in heart surgery].Anaesthesist. 1991 Nov;40(11):602-7. Anaesthesist. 1991. PMID: 1755530 German.
-
Accuracy of methemoglobin detection by pulse CO-oximetry during hypoxia.Anesth Analg. 2010 Jul;111(1):143-8. doi: 10.1213/ANE.0b013e3181c91bb6. Epub 2009 Dec 10. Anesth Analg. 2010. PMID: 20007731
-
Pulse oximetry and finger blood pressure measurement during open-heart surgery.J Clin Monit. 1989 Oct;5(4):221-8. doi: 10.1007/BF01618251. J Clin Monit. 1989. PMID: 2809666
-
The measurement of dyshemoglobins and total hemoglobin by pulse oximetry.Curr Opin Anaesthesiol. 2008 Dec;21(6):805-10. doi: 10.1097/ACO.0b013e328316bb6f. Curr Opin Anaesthesiol. 2008. PMID: 18997533 Review.
-
Retinal oximetry and systemic arterial oxygen levels.Acta Ophthalmol. 2018 Nov;96 Suppl A113:1-44. doi: 10.1111/aos.13932. Acta Ophthalmol. 2018. PMID: 30460761 Review.
Cited by
-
Continuous operation of battery-free implants enables advanced fracture recovery monitoring.Sci Adv. 2025 May 9;11(19):eadt7488. doi: 10.1126/sciadv.adt7488. Epub 2025 May 9. Sci Adv. 2025. PMID: 40344068 Free PMC article.
-
Current state of the art and future directions for implantable sensors in medical technology: Clinical needs and engineering challenges.APL Bioeng. 2023 Sep 27;7(3):031506. doi: 10.1063/5.0152290. eCollection 2023 Sep. APL Bioeng. 2023. PMID: 37781727 Free PMC article. Review.
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Medical