Intraobserver and interobserver reproducibility in the evaluation of ultrasonic pachymetry measurements of central corneal thickness

Affiliations

Intraobserver and interobserver reproducibility in the evaluation of ultrasonic pachymetry measurements of central corneal thickness

S Miglior et al. Br J Ophthalmol. 2004 Feb.

Abstract

Purpose: To assess reproducibility of central corneal thickness (CCT) measurement by means of ultrasonic pachymetry.

Methods: Fifty one volunteers underwent three sessions of CCT measurements, each consisting of three CCT measurements, performed by each of three different observers. Intra- and interobserver reproducibility was calculated by means of intraclass correlation coefficient (ICC). The expected range of variability between two independent evaluations was calculated using scatter plots of each test-retest difference against their mean. The standard deviation of the mean differences in the test-retest scores was used to describe the differences in the score spread.

Results: The ICC ranges of the intra- and interobserver evaluations were 0.95-0.97 and 0.89-0.95 respectively; the expected variability was < or = +/-1% and < or = +/- 2% respectively (95% confidence interval).

Conclusions: The measurement of CCT by means of ultrasonic pachymetry is highly reproducible.

PubMed Disclaimer

References

    1. Carr J, Hersh P. Patient evaluation for refractive surgery. In: Azar DT, ed. Refractive Surgery. Stanford, CT: Appleton & Lange Publishers, 1997:101–9.
    1. Doughty MJ, Zaman ML. Human corneal thickness and its impact on intraocular pressure measures: a review and meta-analysis approach. Surv Ophthalmol 2000;44:367–408. - PubMed
    1. Ehlers N, Hansen FK. Central corneal thickness in low-tension glaucoma. Acta Ophthalmol (Copenh) 1974;52:740–6. - PubMed
    1. Ehlers N, Bramsen T, Sperling S. Applanation tonometry and central corneal thickness. Acta Ophthalmol (Copenh) 1975;53:34–43. - PubMed
    1. Herndon LW, Choudhri SA, Cox T, et al. Central corneal thickness in normal, glaucomatous and ocular hypertensive eyes. Arch Ophthalmol 1975;115:1137–41. - PubMed

Publication types