Comparative depths of cure among various curing light types and methods
- PMID: 14753326
Comparative depths of cure among various curing light types and methods
Abstract
This study evaluated the depth of cure associated with commercial LEDs (light-emitting diodes) (Elipar FreeLight [FL], 3M-ESPE; GC e-Light [EL], GC), high intensity (Elipar TriLight [TL], 3M-ESPE) and very high intensity (Astralis 10 [AS], Ivoclar Vivadent) Quartz Tungsten Halogen (QTH) curing lights. Depth of cure of the various lights/curing modes were compared to a conventional QTH light (Max [Mx], Dentsply-Caulk). Ten exposure regimens were investigated: FL1 - 400 mW/cm2 [40 seconds]; FL2 - 0-400 mW/cm2 [12 seconds] --> 400 mW/cm2 [28 seconds]; EL1 - 750 mW/cm2 [10 pulses x 2 seconds], EL2 - 350 mW/cm2 [40 seconds]; EL3 - 600 mW/cm2 [20 seconds]; EL4 - 0 - 600 mW/cm2 [20 seconds] --> 600 mW/cm2 [20 seconds]; TL1 - 800 mW/cm2 [40 seconds]; TL2 - 100- 800 mW/cm2 [15 seconds] --> 800 mW/cm2 [25 seconds]; AS1 - 1200 mW/cm2 [10 seconds]; MX - 400 mW/cm2 [40 seconds]. Depth of cure was determined by penetration, scraping and microhardness techniques. The results were analyzed using one-way ANOVA/Scheffe's post-hoc test and Pearson's correlation at significance level 0.05 and 0.01, respectively. All light curing regimens met the ISO depth of cure requirement of 1.5 mm with the exception of EL1-EL3 with the microhardness technique. Curing with most modes of EL resulted in significantly lower depths of cure than the control [MX]. No significant difference in depth of cure was observed among the control and the two modes of FL. Curing with TL1 resulted in significantly greater depth of cure compared to MX with all testing techniques. No significant difference in depth of cure was observed between the control and AS1 for all testing techniques except for the penetration technique. The depth of composite cure is light unit and exposure mode dependent. Scraping and penetration techniques were found to correlate well but tend to overestimate depth of cure compared to microhardness.
Similar articles
-
The effectiveness of cure of LED and halogen curing lights at varying cavity depths.Oper Dent. 2003 Nov-Dec;28(6):707-15. Oper Dent. 2003. PMID: 14653284
-
Effectiveness of composite cure associated with different curing modes of LED lights.Oper Dent. 2003 Jul-Aug;28(4):371-7. Oper Dent. 2003. PMID: 12877422
-
Analysis of the degree of conversion of LED and halogen lights using micro-Raman spectroscopy.Oper Dent. 2004 Sep-Oct;29(5):571-7. Oper Dent. 2004. PMID: 15470880
-
Curing lights--the effects of clinical factors on intensity and polymerisation.SADJ. 2002 May;57(5):181-6. SADJ. 2002. PMID: 12174714 Review.
-
Laser curing of dental materials.Dent Clin North Am. 2000 Oct;44(4):923-30. Dent Clin North Am. 2000. PMID: 11048280 Review.
Cited by
-
Effects of Light Attenuation through Dental Tissues on Cure Depth of Composite Resins.Acta Stomatol Croat. 2019 Jun;53(2):95-105. doi: 10.15644/asc53/2/1. Acta Stomatol Croat. 2019. PMID: 31341317 Free PMC article.
-
Effect of Polymerization Time and Home Bleaching Agent on the Microhardness and Surface Roughness of Bulk-Fill Composites: A Scanning Electron Microscopy Study.Scanning. 2019 Jun 2;2019:2307305. doi: 10.1155/2019/2307305. eCollection 2019. Scanning. 2019. PMID: 31281561 Free PMC article.
-
The effects of irradiance and exposure time on the surface roughness of bulk-fill composite resin restorative materials.Saudi Med J. 2018 Feb;39(2):197-202. doi: 10.15537/smj.2018.2.21440. Saudi Med J. 2018. PMID: 29436570 Free PMC article.
-
Contemporary curing profiles: Study of effectiveness of cure and polymerization shrinkage of composite resins: An in vitro study.J Conserv Dent. 2011 Oct;14(4):383-6. doi: 10.4103/0972-0707.87205. J Conserv Dent. 2011. PMID: 22144808 Free PMC article.
-
Microhardness of composite resins at different depths varying the post-irradiation time.J Appl Oral Sci. 2007 Aug;15(4):305-9. doi: 10.1590/s1678-77572007000400012. J Appl Oral Sci. 2007. PMID: 19089149 Free PMC article.
Publication types
MeSH terms
Substances
LinkOut - more resources
Research Materials
Miscellaneous