Novadaq SPY: intraoperative quality assessment in off-pump coronary artery bypass grafting
Abstract
Objectives: Off-pump coronary artery bypass grafting may decrease operative morbidity when compared to on-pump bypass grafting; however, it is technically demanding and thus quality control is essential. This study assesses the clinical feasibility of a new, indocyanine green (ICG)-based imaging system (SPY; Novadaq Technologies; Toronto, ON, Canada) to monitor the quality of anastomoses and grafts in off-pump revascularization.
Setting: Thirty-eight consecutive patients undergoing nonemergent coronary artery bypass grafting without the use of extracorporeal circulation at two Swiss cardiac surgery clinics were included. On completion of bypass grafts, the quality of the grafts was assessed using the ICG-based imaging system. The imaging device comprises an 806-nm laser light source that is used to cause ICG to fluoresce and a near infrared-sensitive charged couple device videocamera that is used to capture the fluorescence images. ICG was administered through the central venous line, and images were acquired during the first pass of the ICG through the field of view. Graft flow (qualitative) and the quality of the grafts and anastomoses were assessed intraoperatively.
Results: Between March 2002 and September 2002, a total of 38 patients (26 men and 12 women; mean +/- SD age, 64.6 +/- 10.5 years; body mass index, 27.1 +/- 2.9) underwent surgery and imaging at two institutions. One hundred seven of 124 grafts (45 arteries and 62 veins) were analyzed. Seventeen grafts could not be assessed due to difficulties in positioning. The imaging system was easy to handle, and no adverse reactions to ICG were observed. Four of the 107 grafts imaged required revision (three anastomotic constrictions and one graft dissection). Each imaging sequence required approximately 1.25 to 2.5 mg of ICG. The images were equivalent to angiography without the need for radiographs and catheter insertion. In addition, the course of coronaries that would otherwise be difficult to locate in obese patients could be detected using the imaging system. Biochemical and ECG data demonstrated an absence of intraoperative or postoperative myocardial damage, and no liver enzyme elevation or renal dysfunction.
Conclusions: This study supports the clinical utility of a ICG-based imaging system for the assessment of the quality of bypass grafts, which appears to be safe and simple to use.
Comment in
-
OPCAB vs CABG: who, what, when, where?Chest. 2004 Mar;125(3):815-6. doi: 10.1378/chest.125.3.815. Chest. 2004. PMID: 15006936 No abstract available.
Similar articles
-
Intraoperative fluorescence imaging system for on-site assessment of off-pump coronary artery bypass graft.JACC Cardiovasc Imaging. 2009 May;2(5):604-12. doi: 10.1016/j.jcmg.2008.12.028. JACC Cardiovasc Imaging. 2009. PMID: 19442948
-
Fractional flow reserve-guided coronary artery bypass grafting: can intraoperative physiologic imaging guide decision making?J Thorac Cardiovasc Surg. 2013 Oct;146(4):824-835.e1. doi: 10.1016/j.jtcvs.2013.06.026. Epub 2013 Aug 1. J Thorac Cardiovasc Surg. 2013. PMID: 23915918
-
Preliminary experience with a novel intraoperative fluorescence imaging technique to evaluate the patency of bypass grafts in total arterial revascularization.Ann Thorac Surg. 2003 Mar;75(3):870-3. doi: 10.1016/s0003-4975(02)04669-6. Ann Thorac Surg. 2003. PMID: 12645709
-
Intraoperative angiography leads to graft revision in coronary artery bypass surgery.Ann Thorac Surg. 2004 Aug;78(2):502-5; discussion 505. doi: 10.1016/j.athoracsur.2004.03.004. Ann Thorac Surg. 2004. PMID: 15276506 Review.
-
Intraoperative coronary graft assessment.Curr Opin Cardiol. 2008 Nov;23(6):568-72. doi: 10.1097/HCO.0b013e3283121cbf. Curr Opin Cardiol. 2008. PMID: 18830071 Review.
Cited by
-
Fluorescently labeled therapeutic antibodies for detection of microscopic melanoma.Laryngoscope. 2013 Nov;123(11):2681-9. doi: 10.1002/lary.24102. Epub 2013 Apr 24. Laryngoscope. 2013. PMID: 23616260 Free PMC article.
-
Molecular fluorescence-guided surgery of peritoneal carcinomatosis of colorectal origin: A narrative review.J Surg Oncol. 2018 Aug;118(2):332-343. doi: 10.1002/jso.25106. Epub 2018 Jun 24. J Surg Oncol. 2018. PMID: 29938400 Free PMC article. Review.
-
Indocyanine green-enhanced fluorescence for assessing parathyroid perfusion during thyroidectomy.Gland Surg. 2016 Oct;5(5):512-521. doi: 10.21037/gs.2016.10.06. Gland Surg. 2016. PMID: 27867866 Free PMC article. Review.
-
Use of panitumumab-IRDye800 to image cutaneous head and neck cancer in mice.Otolaryngol Head Neck Surg. 2013 Jun;148(6):982-90. doi: 10.1177/0194599813482290. Epub 2013 Mar 22. Otolaryngol Head Neck Surg. 2013. PMID: 23525846 Free PMC article.
-
Doxorubicin and indocyanine green loaded superparamagnetic iron oxide nanoparticles with PEGylated phospholipid coating for magnetic resonance with fluorescence imaging and chemotherapy of glioma.Int J Nanomedicine. 2018 Dec 20;14:101-117. doi: 10.2147/IJN.S173954. eCollection 2019. Int J Nanomedicine. 2018. PMID: 30587988 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical
Miscellaneous