The electrode system in impedance-based ventilation measurement
- PMID: 1487276
- DOI: 10.1109/10.168692
The electrode system in impedance-based ventilation measurement
Abstract
In this paper, we determined which electrode types, sizes, and locations were best suited for impedance-based ventilation measurement. Optimal electrodes provide high signal-to-(motion) artifact ratio (SAR) and reliability by meeting the following criteria: 1) low baseline impedance, 2) high adhesion, 3) good physical stability, 4) large effective area, 5) thin with high flexibility. We compared 14 electrodes from two main groups: adhesive-gel and conductive rubber electrodes. Adhesive-gel electrodes are easy to apply, make good body contact, and do not slip during the course of an experiment. We found that higher SAR's are obtained when electrode area is increased by connecting several small electrodes together rather than by using a single electrode with a larger area. The peak SAR is achieved when two electrode arrays (area = 70 cm2) are centered at the 8th intercostal spaces on opposite midaxillary lines. To determine the optimal electrode locations, we placed 32 electrodes on the trunk and recorded impedance between 171 electrode combinations on ten normal adult subjects. Based on these data, we conclude that the SAR's are highest when one electrode is placed on the midpoint between the left and right second intercostal spaces on the sternum and the other electrode is placed in the opposite position on the back.
Similar articles
-
Simultaneous comparison of 1 gel with 4 dry electrode types for electrocardiography.Physiol Meas. 2015 Mar;36(3):513-29. doi: 10.1088/0967-3334/36/3/513. Epub 2015 Feb 16. Physiol Meas. 2015. PMID: 25684219
-
A comparison study of electrodes for neonate electrical impedance tomography.Physiol Meas. 2009 Jun;30(6):S73-84. doi: 10.1088/0967-3334/30/6/S05. Epub 2009 Jun 2. Physiol Meas. 2009. PMID: 19491443
-
Transthoracic impedance study with large self-adhesive electrodes in two conventional positions for defibrillation.Physiol Meas. 2006 Oct;27(10):1009-22. doi: 10.1088/0967-3334/27/10/007. Epub 2006 Aug 25. Physiol Meas. 2006. PMID: 16951460
-
Dry-contact and noncontact biopotential electrodes: methodological review.IEEE Rev Biomed Eng. 2010;3:106-19. doi: 10.1109/RBME.2010.2084078. IEEE Rev Biomed Eng. 2010. PMID: 22275204 Review.
-
Integrated circuits and electrode interfaces for noninvasive physiological monitoring.IEEE Trans Biomed Eng. 2014 May;61(5):1522-37. doi: 10.1109/TBME.2014.2308552. IEEE Trans Biomed Eng. 2014. PMID: 24759282 Review.
Cited by
-
Reduction of motion artefacts during paediatric/infant apnoea monitoring.Med Biol Eng Comput. 1996 Jan;34(1):93-6. doi: 10.1007/BF02637029. Med Biol Eng Comput. 1996. PMID: 8857319 No abstract available.
-
Optimizing bioimpedance measurement configuration for dual-gated nuclear medicine imaging: a sensitivity study.Med Biol Eng Comput. 2011 Jul;49(7):783-91. doi: 10.1007/s11517-011-0787-2. Epub 2011 May 27. Med Biol Eng Comput. 2011. PMID: 21618055
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources
Miscellaneous