Audiometric evaluation of an attempt to optimize the fixation of the transducer of a middle-ear implant to the ossicular chain with bone cement
- PMID: 14961845
- DOI: 10.1111/j.1365-2273.2004.00749.x
Audiometric evaluation of an attempt to optimize the fixation of the transducer of a middle-ear implant to the ossicular chain with bone cement
Abstract
Typically, an implantable hearing device consists of a transducer that is coupled to the ossicular chain and electronics. The coupling is of major importance. The Vibrant Soundbridge (VSB) is such an implantable device; normally, the VSB transducer is fixed to the ossicular chain by means of a special clip that is crimped around the long process of the incus. In addition to crimping, bone cement was used to optimize the fixation in six patients. Long-term results were compared to those of five controls with crimp fixation alone. To assess the effect of bone cement (SerenoCem, Corinthian Medical Ltd, Nottingham, UK) on hearing thresholds, long-term post-surgery thresholds were compared to pre-surgery thresholds. Bone cement did not have any negative effect. Next, to test the hypothesis that aided thresholds might be better with the use of bone cement, aided thresholds were studied. After correction for the severity of hearing loss, only a small difference was found between the two groups at one frequency, viz. 2 kHz. It was concluded that there was no negative effect of using bone cement; however, there is also no reason to use bone cement in VSB users on a regular basis.
Similar articles
-
Our auditory results using the Vibrant Soundbridge on the long process of the incus: 20 years of data.Auris Nasus Larynx. 2018 Feb;45(1):66-72. doi: 10.1016/j.anl.2017.02.007. Epub 2017 Mar 9. Auris Nasus Larynx. 2018. PMID: 28285827
-
Electrocoagulation improving bone cement use in middle-ear surgery: short-term and middle-term results.J Laryngol Otol. 2016 Dec;130(12):1110-1114. doi: 10.1017/S0022215116009014. Epub 2016 Nov 4. J Laryngol Otol. 2016. PMID: 27809939
-
Reinforced active middle ear implant fixation in incus vibroplasty.Ear Hear. 2015 Jan;36(1):72-81. doi: 10.1097/AUD.0000000000000078. Ear Hear. 2015. PMID: 25099400
-
Active middle-ear implant fixation in an unusual place: clinical and audiological outcomes.J Laryngol Otol. 2016 Apr;130(4):404-7. doi: 10.1017/S0022215116000712. J Laryngol Otol. 2016. PMID: 26991877 Review.
-
Application of an active middle ear implant in congenital middle ear malformations: A contemporary review.Braz J Otorhinolaryngol. 2025 May-Jun;91(3):101562. doi: 10.1016/j.bjorl.2025.101562. Epub 2025 Feb 5. Braz J Otorhinolaryngol. 2025. PMID: 39914045 Free PMC article. Review.
Cited by
-
Glass ionomer cement in otological microsurgery: experience over 16 years.Eur Arch Otorhinolaryngol. 2015 Oct;272(10):2749-54. doi: 10.1007/s00405-014-3276-z. Epub 2014 Sep 10. Eur Arch Otorhinolaryngol. 2015. PMID: 25209434
-
[Semi-implantable hearing aids for sensorineural hearing loss and combined hearing loss: experiences at the German Armed Forces Hospital in Ulm].HNO. 2009 Mar;57(3):208-15. doi: 10.1007/s00106-008-1851-6. HNO. 2009. PMID: 19247625 Review. German.
-
[Fully implantable hearing systems].HNO. 2009 Mar;57(3):199-207. doi: 10.1007/s00106-008-1852-5. HNO. 2009. PMID: 19247624 Review. German.
-
[The Vibrant Soundbridge as an active implant in middle ear surgery].HNO. 2014 Jul;62(7):509-19. doi: 10.1007/s00106-014-2884-7. HNO. 2014. PMID: 25056648 Review. German.
-
Intraoperative ABR measurements via the active middle ear implant vibrant soundbridge using different types of couplers.Eur Arch Otorhinolaryngol. 2025 Jul;282(7):3507-3512. doi: 10.1007/s00405-025-09248-5. Epub 2025 Feb 1. Eur Arch Otorhinolaryngol. 2025. PMID: 39891691 Free PMC article.
MeSH terms