Automated detection of five human herpes virus DNAs by a set of LightCycler PCRs complemented with a single multiple internal control
- PMID: 14962786
- DOI: 10.1016/S1386-6532(03)00121-5
Automated detection of five human herpes virus DNAs by a set of LightCycler PCRs complemented with a single multiple internal control
Abstract
Background: Herpes viruses represent important causes of morbidity and mortality especially in immuno-compromised patients. To assist in rapid diagnosis real-time PCR assays have been developed for the detection of herpes virus DNA in patient specimens. A recently described set of real-time PCR assays using LightCycler technology enabled parallel detection of DNA from cytomegalovirus (CMV), Epstein-Barr virus (EBV), herpes simplex virus type 1 and 2 (HSV-1/-2), and varicella-zoster virus (VZV) by using a single LightCycler program [J. Clin. Virol. 26 (2003) 85]. The set of assays lacked automation of DNA purification and of PCR mixture preparation, and was not furnished with measures to monitor for sample adequacy.
Objectives: Development of a set of automated LightCycler-PCR assays for the detection CMV-, EBV-, HSV-1/-2- and VZV-DNA in plasma samples and complementation of the assays with internal amplification controls (ICs).
Study design: The MagNA Pure LC instrument was used for automated DNA purification and automated preparation of PCR mixtures. A single multiple IC-DNA specific for all four herpes virus type-specific PCRs was generated and used in all four LightCycler assays. Detection limits were determined and clinical samples were evaluated.
Results: With quantified herpes virus type-specific reference DNA spiked into EDTA plasma, the detection limits were found at 250 copies/ml of CMV-, EBV-, HSV-1/-2-DNA and at 500 copies/ml of VZV-DNA. The novel set of assays was evaluated by testing 112 EDTA plasma samples. The use of the IC led to the detection of PCR-inhibited samples.
Conclusion: The set of automated LightCycler assays was found rapid, markedly labour saving and suitable for the routine diagnostic laboratory. The use of the one internal control molecule simplified the assay protocol and allowed monitoring for sample adequacy.
Similar articles
-
Routine use of a highly automated and internally controlled real-time PCR assay for the diagnosis of herpes simplex and varicella-zoster virus infections.J Clin Virol. 2004 May;30(1):39-44. doi: 10.1016/j.jcv.2003.08.006. J Clin Virol. 2004. PMID: 15072752
-
Rapid quantitative PCR assays for the simultaneous detection of herpes simplex virus, varicella zoster virus, cytomegalovirus, Epstein-Barr virus, and human herpesvirus 6 DNA in blood and other clinical specimens.J Med Virol. 2008 Mar;80(3):467-77. doi: 10.1002/jmv.21095. J Med Virol. 2008. PMID: 18205230
-
Parallel detection of five human herpes virus DNAs by a set of real-time polymerase chain reactions in a single run.J Clin Virol. 2003 Jan;26(1):85-93. doi: 10.1016/s1386-6532(02)00042-2. J Clin Virol. 2003. PMID: 12589838
-
Herpes viruses in transplant recipients: HSV, VZV, human herpes viruses, and EBV.Infect Dis Clin North Am. 2010 Jun;24(2):373-93. doi: 10.1016/j.idc.2010.01.003. Infect Dis Clin North Am. 2010. PMID: 20466275 Review.
-
Progress in diagnosing herpesvirus infections.Nagoya J Med Sci. 1999 Nov;62(3-4):83-97. Nagoya J Med Sci. 1999. PMID: 10689870 Review.
Cited by
-
Prediction of cytomegalovirus (CMV) plasma load from evaluation of CMV whole-blood load in samples from renal transplant recipients.J Clin Microbiol. 2008 Feb;46(2):493-8. doi: 10.1128/JCM.01499-07. Epub 2007 Dec 5. J Clin Microbiol. 2008. PMID: 18057128 Free PMC article.
-
An Internal Reference Control Duplex Real-Time Polymerase Chain Reaction Assay for Detecting Bacterial Contamination in Blood Products.PLoS One. 2015 Jul 31;10(7):e0134743. doi: 10.1371/journal.pone.0134743. eCollection 2015. PLoS One. 2015. PMID: 26230627 Free PMC article.
-
Application of molecular diagnostic techniques for viral testing.Open Virol J. 2012;6:104-14. doi: 10.2174/1874357901206010104. Epub 2012 Nov 30. Open Virol J. 2012. PMID: 23248732 Free PMC article.
-
Development and utility of an internal threshold control (ITC) real-time PCR assay for exogenous DNA detection.PLoS One. 2012;7(5):e36461. doi: 10.1371/journal.pone.0036461. Epub 2012 May 3. PLoS One. 2012. PMID: 22570718 Free PMC article.
-
Development and Validation of a Laboratory-Developed Multiplex Real-Time PCR Assay on the BD Max System for Detection of Herpes Simplex Virus and Varicella-Zoster Virus DNA in Various Clinical Specimens.J Clin Microbiol. 2015 Jun;53(6):1921-6. doi: 10.1128/JCM.03692-14. Epub 2015 Apr 15. J Clin Microbiol. 2015. PMID: 25878344 Free PMC article.
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Other Literature Sources