Clinical Trial
. 2004 Mar 1;93(5):633-6.
doi: 10.1016/j.amjcard.2003.11.037.

Treatment of very small vessels with 2.25-mm diameter sirolimus-eluting stents (from the RESEARCH registry)

Affiliations
Clinical Trial

Treatment of very small vessels with 2.25-mm diameter sirolimus-eluting stents (from the RESEARCH registry)

Pedro A Lemos et al. Am J Cardiol. .

Abstract

A total of 91 patients with 112 lesions received 2.25-mm sirolimus-eluting stents (SESs), and these lesions were compared with those treated with SESs of > or =2.5-mm diameter in the same procedure (n = 109). The reference diameters were 1.88 +/- 0.34 and 2.52 +/- 0.57 mm, respectively (p <0.01). At follow-up, the late lumen loss was 0.07 +/- 0.48 mm for the 2.25-mm SES versus 0.03 +/- 0.38 mm for the larger SES (p = 0.5), and the binary restenosis rate was 10.7% versus 3.9%, respectively (p = 0.1). The 12-month target lesion revascularization rate was 5.5%. In conclusion, 2.25-mm SESs were associated with low rates of clinical and angiographic late complications.