Application of radiofrequency energy in surgical and interventional procedures: are there interactions with ICDs?

Affiliations

Application of radiofrequency energy in surgical and interventional procedures: are there interactions with ICDs?

Michael Fiek et al. Pacing Clin Electrophysiol. 2004 Mar.

Abstract

During surgical and interventional procedures, interference may occur between ICDs and electrical cautery or with the application of RF energy. This may lead to the false induction of ICD therapies or could even result in device malfunction, which represents a potential perioperative hazard for the patient. This study analyzed the intraoperative interactions in 45 consecutive ICD patients in reference to different surgical and interventional procedures. A total of 33 surgical operations (general surgery [n = 14], urologic [n = 5], abdominal [n = 10], gynecological [n = 2], thoracic [n = 1], neurosurgical [n = 1]) and 12 interventional therapies (RF catheter ablation [n = 10], endoscopic papillotomy [n = 2]) were performed. The ICD devices were all located in left pectoral position and consisted of 25 single and 20 dual chamber defibrillators. During the procedure, tachyarrhythmia detection (VF 296 +/- 20 ms, VT 376 +/- 49 ms) of the devices was maintained active (monitoring mode), only ICD therapies were inactivated. The indifferent electrode of the electrical cauter/RF generator was placed in standard positions (right/left mid-femoral position [n = 27/8], thoracic spine area [n = 10]). After the procedure, the ICD memory was checked for detections and for changes in the programming. There was no oversensing, reprogramming, or damage of any defibrillator caused by RF energy. Despite the lack of undesired interactions, ICDs should be inactivated preoperatively to assure maximum patient safety. However, should inactivation not be possible, or the achievement uncertain, electromagnetic interference is highly unlikely.