Real-time continuous intraoperative electromyographic and somatosensory evoked potential recordings in spinal surgery: correlation of clinical and electrophysiologic findings in a prospective, consecutive series of 213 cases
- PMID: 15014279
- DOI: 10.1097/01.brs.0000115144.30607.e9
Real-time continuous intraoperative electromyographic and somatosensory evoked potential recordings in spinal surgery: correlation of clinical and electrophysiologic findings in a prospective, consecutive series of 213 cases
Abstract
Study design: Retrospective analysis of a prospectively accrued series of 213 consecutive patients who underwent intraoperative neurophysiologic monitoring with electromyography and somatosensory-evoked potentials during thoracolumbar spine surgery.
Objectives: To study the incidence of significant intraoperative electrophysiologic changes and new postoperative neurologic deficits.
Summary of background data: Continuous intraoperative electromyography and somatosensory-evoked potentials are frequently used in spinal surgery to prevent neural injury. However, only limited data are available on the sensitivity, specificity, and predictive values of intraoperative electrophysiologic changes with regard to the occurrence of new postoperative neurologic deficits.
Methods: We examined data on patients who underwent intraoperative monitoring with continuous lower limb electromyography and somatosensory-evoked potentials. The analysis focused on the correlation of intraoperative electrophysiologic changes with the development of new neurologic deficits.
Results: A total of 213 patients underwent surgery on a total of 378 levels; 32.4% underwent an instrumented fusion. Significant electromyograph activation was observed in 77.5% of the patients and significant somatosensory-evoked potential changes in 6.6%. Fourteen patients (6.6%) had new postoperative neurologic symptoms. Of those, all had significant electromyograph activation, but only 4 had significant somatosensory-evoked potential changes. Intraoperative electromyograph activation had a sensitivity of 100% and a specificity of 23.7% for the detection of a new postoperative neurologic deficit. Somatosensory-evoked potentials had a sensitivity of 28.6% and specificity of 94.7%.
Conclusions: Intraoperative electromyographic activation has a high sensitivity for the detection of a newpostoperative neurologic deficit but a low specificity. In contrast, somatosensory-evoked potentials have low sensitivity but high specificity. Combined intraoperative neurophysiologic monitoring with electromyography and somatosensory-evoked potentials is helpful for predicting and possibly preventing neurologic injury during thoracolumbar spine surgery.
Similar articles
-
Multimodality intraoperative neurophysiologic monitoring findings during surgery for adult tethered cord syndrome: analysis of a series of 44 patients with long-term follow-up.Spine (Phila Pa 1976). 2006 Aug 15;31(18):2095-102. doi: 10.1097/01.brs.0000231687.02271.b6. Spine (Phila Pa 1976). 2006. PMID: 16915095
-
Continuous electromyographic monitoring to detect nerve root injury during thoracolumbar scoliosis surgery.Spine (Phila Pa 1976). 1997 Nov 1;22(21):2547-50. doi: 10.1097/00007632-199711010-00016. Spine (Phila Pa 1976). 1997. PMID: 9383863
-
Intraoperative neurophysiologic spinal cord monitoring in thoracolumbar burst fractures.Spine (Phila Pa 1976). 2009 Nov 15;34(24):2662-8. doi: 10.1097/BRS.0b013e3181bf151b. Spine (Phila Pa 1976). 2009. PMID: 19910769
-
What is the predictive value of intraoperative somatosensory evoked potential monitoring for postoperative neurological deficit in cervical spine surgery?-a meta-analysis.Spine J. 2021 Apr;21(4):555-570. doi: 10.1016/j.spinee.2021.01.010. Epub 2021 Jan 16. Spine J. 2021. PMID: 33460808 Review.
-
Intraoperative neurophysiologic monitoring during spinal surgery.J Am Acad Orthop Surg. 2007 Sep;15(9):549-60. doi: 10.5435/00124635-200709000-00005. J Am Acad Orthop Surg. 2007. PMID: 17761611 Review.
Cited by
-
Relationship between Spinal Cord Volume and Spinal Cord Injury due to Spinal Shortening.PLoS One. 2015 May 22;10(5):e0127624. doi: 10.1371/journal.pone.0127624. eCollection 2015. PLoS One. 2015. PMID: 26001196 Free PMC article.
-
Intraoperative electrophysiological monitoring during posterior craniocervical distraction and realignment for congenital craniocervical anomaly.Eur Spine J. 2015 Apr;24(4):671-8. doi: 10.1007/s00586-015-3791-7. Epub 2015 Feb 7. Eur Spine J. 2015. PMID: 25663645
-
Intraoperative Neurophysiological Monitoring for Intradural Extramedullary Spinal Tumours.Global Spine J. 2024 May;14(4):1304-1315. doi: 10.1177/21925682221139822. Epub 2022 Nov 21. Global Spine J. 2024. PMID: 36411068 Free PMC article.
-
Should evoked potential monitoring be used in degenerative cervical spine surgery? A systematic review.J Orthop Traumatol. 2019 Apr 2;20(1):19. doi: 10.1186/s10195-019-0524-4. J Orthop Traumatol. 2019. PMID: 30941518 Free PMC article.
-
Intraoperative Neurophysiological Monitoring : A Review of Techniques Used for Brain Tumor Surgery in Children.J Korean Neurosurg Soc. 2018 May;61(3):363-375. doi: 10.3340/jkns.2018.0078. Epub 2018 May 1. J Korean Neurosurg Soc. 2018. PMID: 29742889 Free PMC article. Review.
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical