Initial experience with remote catheter ablation using a novel magnetic navigation system: magnetic remote catheter ablation

Affiliations

Initial experience with remote catheter ablation using a novel magnetic navigation system: magnetic remote catheter ablation

Sabine Ernst et al. Circulation. .

Abstract

Background: Catheters are typically stiff and incorporate a pull-wire mechanism to allow tip deflection. While standing at the patient's side, the operator manually navigates the catheter in the heart using fluoroscopic guidance.

Methods and results: A total of 42 patients (32 female; mean age, 55+/-15 years) underwent ablation of common-type (slow/fast) or uncommon-type (slow/slow) atrioventricular nodal reentrant tachycardia (AVNRT) with the use of the magnetic navigation system Niobe (Stereotaxis, Inc). It consists of 2 computer-controlled permanent magnets located on opposite sides of the patient, which create a steerable external magnetic field (0.08 T). A small magnet embedded in the catheter tip causes the catheter to align and to be steered by the external magnetic field. A motor drive advances or retracts the catheter, enabling complete remote navigation. Radiofrequency current was applied with the use of a remote-controlled 4-mm, solid-tip, magnetic navigation-enabled catheter (55 degrees C, maximum 40 W, 60 seconds) in all patients. The investigators, who were situated in the control room, performed the ablation using a mean of 7.2+/-4.7 radiofrequency current applications (mean fluoroscopy time, 8.9+/-6.2 minutes; procedure duration, 145+/-43 minutes). Slow pathway ablation was achieved in 15 patients, whereas slow pathway modulation was the end point in the remaining patients. There were no complications.

Conclusions: The Niobe magnetic navigation system is a new platform technology allowing remote-controlled navigation of an ablation catheter. In conjunction with a motor drive unit, this system was used successfully to perform completely remote-controlled mapping and ablation in patients with AVNRT.