Purpose: To determine predictors of patient radiation exposure due to percutaneous coronary interventions (PCI) and to compare our results with the "preliminary reference levels", recently proposed by the European DIMOND research cardiology group: i. e., 75 Gy.cm(2) for dose area kerma product (DAP), 17 min for fluoroscopy time (T(F)), and 1,300 for cinegraphic frames (F).
Material and methods: For 642 PCI-exclusive of the fraction for diagnostic catheterization to avoid statistical confounder effects-we measured total DAP, cinegraphic (DAP(C)) and fluoroscopic (DAP(F)) fractions, the number of cinegraphic frames and runs, and T(F). DAP(C)/F and DAP(F)/s were calculated to indicate the quality of focusing to the region of interest.
Results: The mean total patient DAP for elective one-, two-, and three-vessel PCI amounted to 6.7, 11.6, and 19.4 Gy.cm(2), for PCI of focal in-stent restenoses to 4.2 Gy.cm(2), and for excimer laser angioplasty of diffuse in-stent restenoses to 19.4 Gy.cm(2), respectively. Recanalization of chronic occlusions and PCI in acute myocardial infarction occasioned mean levels of 16.0 and 17.3 Gy.cm(2). Implantation of one and > or = two stents during one-vessel PCI significantly increased total mean DAP from a baseline level of 5.7 up to 7.1 and to 13.8 Gy.cm(2). DAP significantly varied according to the various PCI target regions and amounted to 4.0, 4.5, and 5.5 Gy.cm(2) for intermedius, diagonal, and left anterior descending arteries, to 4.9, 5.0, and 7.0 Gy.cm(2) for obtuse marginal, left posterolateral, and circumflex arteries, to 8.3, 9.1, and 9.5 Gy.cm(2) for proximal/mid right coronary segments, posterior descending, and right posterolateral arteries, and to 11.6 Gy.cm(2) for saphenous vein grafts, respectively.
Conclusion: This study, carried through by consistent use of radiation-reducing techniques, enables a reliable scoring of patient radiation exposure according to complexity and target vessel of the intended PCI. Our 95th percentiles for elective PCI, for recanalizations of chronic occlusions, and for emergency PCI advise reference levels of 22, 32, and 42 Gy.cm(2) for DAP, of 16, 25, and 24 min for T(F), and of 400, 600, and 700 cinegraphic frames, respectively.
Similar articles
-
Effective techniques for reduction of radiation dosage to patients undergoing invasive cardiac procedures.Br J Radiol. 2003 Jun;76(906):406-13. doi: 10.1259/bjr/82051842. Br J Radiol. 2003. PMID: 12814927
-
[Fluoroscopy time -- an overestimated factor for patient radiation exposure in invasive cardiology].Rofo. 2005 Jun;177(6):812-7. doi: 10.1055/s-2005-858189. Rofo. 2005. PMID: 15902630 German.
-
Predialing the number of cinegraphic frames enables an effective patient dose due to coronary angiography of 0.8 mSv.Rofo. 2003 Dec;175(12):1706-10. doi: 10.1055/s-2003-45341. Rofo. 2003. PMID: 14661143
-
Myocardial infarction induced by myocardial bridging: report of successful percutaneous therapy and review of the literature.J Invasive Cardiol. 2005 Dec;17(12):E52-4. J Invasive Cardiol. 2005. PMID: 16327041 Review.
-
[Excimer laser assisted angioplasty. Immediate results in the treatment of complex coronary lesions].Arch Mal Coeur Vaiss. 1996 Apr;89(4):407-15. Arch Mal Coeur Vaiss. 1996. PMID: 8762999 Review. French.
Cited by
-
Patient radiation doses in interventional cardiology procedures.Curr Cardiol Rev. 2009 Jan;5(1):1-11. doi: 10.2174/157340309787048059. Curr Cardiol Rev. 2009. PMID: 20066141 Free PMC article.
-
Reference levels and patient doses in interventional cardiology procedures in Greece.Eur Radiol. 2013 Aug;23(8):2324-32. doi: 10.1007/s00330-013-2813-2. Epub 2013 Apr 6. Eur Radiol. 2013. PMID: 23559142
-
Predictors of Excess Patient Radiation Exposure During Chronic Total Occlusion Coronary Intervention: Insights From a Contemporary Multicentre Registry.Can J Cardiol. 2017 Apr;33(4):478-484. doi: 10.1016/j.cjca.2016.11.002. Epub 2016 Nov 11. Can J Cardiol. 2017. PMID: 28169091 Free PMC article.
-
Status of radiation protection in various interventional cardiology procedures in the Asia Pacific region.Heart Asia. 2011 Jan 1;3(1):16-24. doi: 10.1136/ha.2009.001180. eCollection 2011. Heart Asia. 2011. PMID: 27325974 Free PMC article.
-
Radiation reduction during percutaneous coronary intervention: A new protocol with a low frame rate and selective fluoroscopic image storage.Medicine (Baltimore). 2017 Jul;96(30):e7517. doi: 10.1097/MD.0000000000007517. Medicine (Baltimore). 2017. PMID: 28746198 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Medical
Miscellaneous