Comparative Study
. 2004 Sep;14(9):1590-5.
doi: 10.1007/s00330-004-2316-2. Epub 2004 Apr 1.

3D imaging with an isocentric mobile C-arm comparison of image quality with spiral CT

Affiliations
Comparative Study

3D imaging with an isocentric mobile C-arm comparison of image quality with spiral CT

Dorothea Kotsianos et al. Eur Radiol. 2004 Sep.

Abstract

The purpose of this study was to evaluate the image quality of the new 3D imaging system (ISO-C-3D) for osteosyntheses of tibial condylar fractures in comparison with spiral CT (CT). Sixteen human cadaveric knees were examined with a C-arm 3D imaging system and spiral computed tomography. Various screws and plates of steel and titanium were used for osteosynthesis in these specimens. Image quality and clinical value of multiplanar (MP) reformatting of both methods were analyzed. In addition, five patients with tibial condylar fractures were examined for diagnosis and intra-operative control. The image quality of the C-arm 3D imaging system in the cadaveric study was rated as significantly worse than that of spiral CT with and without prostheses. After implantation of prostheses an increased incidence of artifacts was observed, but the diagnostic accuracy was not affected. Titanium implants caused the smallest number of artifacts. The image quality of ISO-C is inferior to CT, and metal artifacts were more prominent, but the clinical value was equal. ISO-C-3D can be useful in planning operative reconstructions and can verify the reconstruction of articular surfaces and the position of implants with diagnostic image quality.

PubMed Disclaimer

References

    1. Orthopedics. 1995 Jul;18(7):623-6 - PubMed
    1. Invest Radiol. 1993 Feb;28(2):146-9 - PubMed
    1. Swiss Surg. 1998;(4):180-6 - PubMed
    1. J Comput Assist Tomogr. 2000 Jan-Feb;24(1):165-72 - PubMed
    1. Orthopedics. 2000 Jul;23 (7):713-5 - PubMed

Publication types

MeSH terms

LinkOut - more resources