Comparison of complementary metal oxide semiconductor and charge-coupled device intraoral X-ray detectors using subjective image quality
- PMID: 15070845
- DOI: 10.1259/dmfr/19990417
Comparison of complementary metal oxide semiconductor and charge-coupled device intraoral X-ray detectors using subjective image quality
Abstract
Objective: To compare the subjective image quality of the newer generation Schick CDR detector employing complementary metal oxide semiconductor (CMOS) technology with images using the earlier generation charge-coupled device (CCD) Schick CDR detector.
Methods: All radiographic images were made using the same formalin-fixed adult cadaver maxilla with surrounding natural soft tissues in place. The X-ray generator used was a Villa Sistemi Medicali Diamatic srl AP/Explor X operated at 70 kVp and 8 mA. The source-to-detector distance was set at 38 cm and an optical bench was used to ensure reproducible beam geometry. A range of exposures was applied for both detectors. A panel of nine dentists independently observed and evaluated images made at each exposure. For both detectors, the three images ranked highest were randomized for re-evaluation in panels of six images. Each image was repeated randomly a total of 10 times. Features chosen as observation points were: (1) proximal dental caries; (2) gingival soft tissues; (3) cortical bone; (4) root canal space; (5) root apices; (6) periodontal ligament space; and (7) endodontic instrument tip clarity. Comparisons were made by use of odds ratio analysis applying a 95% confidence level. Interrater and intrarater reliabilities were computed to assess consistency in observer ratings.
Results: The CMOS sensor was rated as outperforming its CCD predecessor for depiction of cortical bone and root apices; the CCD detector was only rated superior for depiction of root canal space. No significant difference was found between the two detectors in perceived depiction of proximal dental caries, gingival soft tissues, periodontal ligament space or endodontic instruments. Combining rating scores from each of the tasks, CMOS and CCD detectors had a similar proportion of image ratings of excellent, acceptable and poor.
Conclusions: Regarding subjective image quality, the Schick CMOS and CCD detectors were perceived to produce radiographic images of similar overall quality.
Similar articles
-
Comparison of three intra-oral storage phosphor systems using subjective image quality.Dentomaxillofac Radiol. 2000 Sep;29(5):272-6. doi: 10.1038/sj/dmfr/4600532. Dentomaxillofac Radiol. 2000. PMID: 10980561
-
Comparison of observer performance in determining the position of endodontic files with physical measures in the evaluation of dental X-ray imaging systems.Dentomaxillofac Radiol. 2000 Jul;29(4):216-22. doi: 10.1038/sj/dmfr/4600533. Dentomaxillofac Radiol. 2000. PMID: 10918454
-
Radiographic endodontic working length estimation: comparison of three digital image receptors.Oral Surg Oral Med Oral Pathol Oral Radiol Endod. 2008 Oct;106(4):604-8. doi: 10.1016/j.tripleo.2008.04.031. Epub 2008 Aug 20. Oral Surg Oral Med Oral Pathol Oral Radiol Endod. 2008. PMID: 18718798
-
Digital imaging for dental caries.Dent Clin North Am. 2000 Apr;44(2):319-38, vi. Dent Clin North Am. 2000. PMID: 10740771 Review.
-
Panoramic radiography in dental diagnostics.Swed Dent J Suppl. 1996;119:1-26. Swed Dent J Suppl. 1996. PMID: 8971997 Review.
Cited by
-
Effect of ambient light level at the monitor surface on digital radiographic evaluation of approximal carious lesions: an in vitro study.Dentomaxillofac Radiol. 2012 Mar;41(3):192-6. doi: 10.1259/dmfr/15422221. Dentomaxillofac Radiol. 2012. PMID: 22378753 Free PMC article.
-
Characteristics of radiographic images acquired with CdTe, CCD and CMOS detectors in skull radiography.Imaging Sci Dent. 2020 Dec;50(4):339-346. doi: 10.5624/isd.2020.50.4.339. Epub 2020 Dec 15. Imaging Sci Dent. 2020. PMID: 33409143 Free PMC article.
-
Artificial Intelligence through Wireless Sensors Applied in Restorative Dentistry: A Systematic Review.Dent J (Basel). 2024 Apr 24;12(5):120. doi: 10.3390/dj12050120. Dent J (Basel). 2024. PMID: 38786518 Free PMC article. Review.
-
Diagnostic accuracy of different imaging modalities in detection of proximal caries.Dentomaxillofac Radiol. 2010 Dec;39(8):501-11. doi: 10.1259/dmfr/28628723. Dentomaxillofac Radiol. 2010. PMID: 21062944 Free PMC article.
-
Performance of Oral Cavity Sensors: A Systematic Review.Sensors (Basel). 2023 Jan 4;23(2):588. doi: 10.3390/s23020588. Sensors (Basel). 2023. PMID: 36679385 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Miscellaneous