Assessment of radiation safety instructions to patients based on measured dose rates following prostate brachytherapy
- PMID: 15110306
- DOI: 10.1016/j.brachy.2004.02.002
Assessment of radiation safety instructions to patients based on measured dose rates following prostate brachytherapy
Abstract
Purpose: To validate radiation safety instructions to patients and to evaluate the potential radiation doses to members of the public after (125)I or (103)Pd prostate implantation.
Methods and materials: Radiation dose rate measurements were made in the immediate postoperative period on 636 consecutive patients with stage T1-T2 prostate cancer who underwent transperineal (125)I or (103)Pd implantation at Memorial Sloan-Kettering Cancer Center during the period from August 1995 through January 2003.
Results: The mean radiation dose rate at the anterior skin surface following a prostate implant was 37 microSv/hr for (125)I and 8 microSv/hr for (103)Pd. At 30 cm from the anterior skin surface, these dose rates were reduced to 6 microSv/hr for (125)I and 3 microSv/hr for (103)Pd. At 1 m from the anterior skin surface the dose rates from both types of implants were reduced to less than 1 microSv/hr. The effect of body weight on dose rates from (125)I sources was examined for a select sub-group of patients and the measured dose rate was found to decrease with increasing body weight. In another group of patients, dose rate measurements were made on both lateral skin surfaces and were less than 16.8 microSv/hr in all cases. Assuming a 33% occupancy factor and utilizing the mean measured dose rate for (125)I, the time required to reach an effective dose equivalent limit of 5 mSv for caregivers was estimated to be 19 days on contact with the skin surface. Using a similar calculation, the lifetime doses for (125)I at a distance of 30 cm from the anterior skin surface, as well as the lifetime doses for (103)Pd on contact with the skin surface and at 30 cm from the anterior skin surface can be shown to be less than 5 mSv.
Conclusions: The large number of cases available for this study permits a validation of radiation safety recommendations and provides concrete information from which the permitted exposure times following implantation can be estimated. The data support the conclusion that patients treated with these implants do not represent a radiation risk to members of the public.
Similar articles
-
Radiation exposure after permanent prostate brachytherapy.Radiother Oncol. 2006 Apr;79(1):65-9. doi: 10.1016/j.radonc.2006.02.010. Epub 2006 Mar 27. Radiother Oncol. 2006. PMID: 16564593
-
Radiation safety aspects of brachytherapy for prostate cancer using permanently implanted sources. A report of ICRP Publication 98.Ann ICRP. 2005;35(3):iii-vi, 3-50. doi: 10.1016/j.icrp.2005.07.001. Ann ICRP. 2005. PMID: 16330284 Review.
-
Radiation exposure to general public after permanent brachytherapy for prostate cancer.Radiat Prot Dosimetry. 2011 Jul;146(1-3):229-30. doi: 10.1093/rpd/ncr156. Epub 2011 Apr 20. Radiat Prot Dosimetry. 2011. PMID: 21511724
-
[Physical and technical quality assurance and radiation protection in transperineal interstitial permanent prostate brachytherapy with 125-iodine seeds].Strahlenther Onkol. 2002 Dec;178(12):667-75. doi: 10.1007/s00066-002-0997-5. Strahlenther Onkol. 2002. PMID: 12491054 German.
-
Urinary morbidity and incontinence following transurethral resection of the prostate after brachytherapy.J Urol. 2005 Mar;173(3):808-12. doi: 10.1097/01.ju.0000152698.20487.0e. J Urol. 2005. PMID: 15711274 Review.
Cited by
-
Less-restrictive, patient-specific radiation safety precautions can be safely prescribed after permanent seed implantation.Brachytherapy. 2010 Apr-Jun;9(2):101-11. doi: 10.1016/j.brachy.2009.06.006. Epub 2009 Oct 22. Brachytherapy. 2010. PMID: 19853532 Free PMC article.
-
A novel perineal shield for low-dose-rate prostate brachytherapy.J Contemp Brachytherapy. 2015 Jun;7(3):197-202. doi: 10.5114/jcb.2015.52071. Epub 2015 Jun 8. J Contemp Brachytherapy. 2015. PMID: 26207107 Free PMC article.
-
Radiation protection of persons living close to patients with radioactive implants.Strahlenther Onkol. 2010 Feb;186(2):107-112. doi: 10.1007/s00066-010-2073-x. Epub 2010 Jan 28. Strahlenther Onkol. 2010. PMID: 20127228
-
Radiation safety of receptive anal intercourse with prostate cancer patients treated with low-dose-rate brachytherapy.Brachytherapy. 2016 Jul-Aug;15(4):420-425. doi: 10.1016/j.brachy.2016.03.012. Epub 2016 May 12. Brachytherapy. 2016. PMID: 27180125 Free PMC article.
-