The use of an aSi-based EPID for routine absolute dosimetric pre-treatment verification of dynamic IMRT fields
- PMID: 15110457
- DOI: 10.1016/j.radonc.2004.02.018
The use of an aSi-based EPID for routine absolute dosimetric pre-treatment verification of dynamic IMRT fields
Abstract
Background and purpose: In parallel with the increased use of intensity modulated radiation treatment (IMRT) fields in radiation therapy, flat panel amorphous silicon (aSi) detectors are becoming the standard for online portal imaging at the linear accelerator. In order to minimise the workload related to the quality assurance of the IMRT fields, we have explored the possibility of using a commercially available aSi portal imager for absolute dosimetric verification of the delivery of dynamic IMRT fields.
Patients and methods: We investigated the basic dosimetric characteristics of an aSi portal imager (aS500, Varian Medical Systems), using an acquisition mode especially developed for portal dose (PD) integration during delivery of a-static or dynamic-radiation field. Secondly, the dose calculation algorithm of a commercially available treatment planning system (Cadplan, Varian Medical Systems) was modified to allow prediction of the PD image, i.e. to compare the intended fluence distribution with the fluence distribution as actually delivered by the dynamic multileaf collimator. Absolute rather than relative dose prediction was applied. The PD image prediction was compared to the corresponding acquisition for several clinical IMRT fields by means of the gamma evaluation method.
Results and conclusions: The acquisition mode is accurate in integrating all PD over a wide range of monitor units, provided detector saturation is avoided. Although the dose deposition behaviour in the portal image detector is not equivalent to the dose to water measurements, it is reproducible and self-consistent, lending itself to quality assurance measurements. Gamma evaluations of the predicted versus measured PD distribution were within the pre-defined acceptance criteria for all clinical IMRT fields, i.e. allowing a dose difference of 3% of the local field dose in combination with a distance to agreement of 3 mm.
Similar articles
-
A simple theoretical verification of monitor unit calculation for intensity modulated beams using dynamic mini-multileaf collimation.Radiother Oncol. 2004 May;71(2):235-41. doi: 10.1016/j.radonc.2004.02.014. Radiother Oncol. 2004. PMID: 15110458
-
[Application of amorphous silicon electronic portal imaging device (a-Si EPID) to dosimetry quality assurance of radiation therapy].Ai Zheng. 2007 Nov;26(11):1272-5. Ai Zheng. 2007. PMID: 17991332 Chinese.
-
A literature review of electronic portal imaging for radiotherapy dosimetry.Radiother Oncol. 2008 Sep;88(3):289-309. doi: 10.1016/j.radonc.2008.07.008. Epub 2008 Aug 14. Radiother Oncol. 2008. PMID: 18706727 Review.
-
SIFT: a method to verify the IMRT fluence delivered during patient treatment using an electronic portal imaging device.Int J Radiat Oncol Biol Phys. 2004 Nov 1;60(3):981-93. doi: 10.1016/j.ijrobp.2004.06.025. Int J Radiat Oncol Biol Phys. 2004. PMID: 15465217
-
Quality assurance of intensity-modulated radiotherapy.Semin Radiat Oncol. 2002 Jul;12(3):219-28. doi: 10.1053/srao.2002.33700. Semin Radiat Oncol. 2002. PMID: 12118387 Review.
Cited by
-
Comparison between an in-house 1D profile correction method and a 2D correction provided in Varian's PDPC Package for improving the accuracy of portal dosimetry images.J Appl Clin Med Phys. 2015 Mar 8;16(2):4973. doi: 10.1120/jacmp.v16i2.4973. J Appl Clin Med Phys. 2015. PMID: 26103173 Free PMC article.
-
Beam perturbation characteristics of a 2D transmission silicon diode array, Magic Plate.J Appl Clin Med Phys. 2016 Mar 8;17(2):85-98. doi: 10.1120/jacmp.v17i2.5932. J Appl Clin Med Phys. 2016. PMID: 27074475 Free PMC article.
-
Comparison of measured and computed portal dose for IMRT treatment.J Appl Clin Med Phys. 2006 Aug 24;7(3):65-79. doi: 10.1120/jacmp.v7i3.2281. J Appl Clin Med Phys. 2006. PMID: 17533343 Free PMC article.
-
A method for in vivo treatment verification of IMRT and VMAT based on electronic portal imaging device.Radiat Oncol. 2021 Dec 4;16(1):232. doi: 10.1186/s13014-021-01953-9. Radiat Oncol. 2021. PMID: 34863229 Free PMC article.
-
A multi-institutional evaluation of machine performance check system on treatment beam output and symmetry using statistical process control.J Appl Clin Med Phys. 2019 Mar;20(3):71-80. doi: 10.1002/acm2.12547. Epub 2019 Feb 20. J Appl Clin Med Phys. 2019. PMID: 30786139 Free PMC article.