Initial experience with miniature axial flow ventricular assist devices for postcardiotomy heart failure
- PMID: 15111158
- DOI: 10.1016/j.athoracsur.2003.10.013
Initial experience with miniature axial flow ventricular assist devices for postcardiotomy heart failure
Abstract
Background: The recently introduced Impella Recover (Impella CardioSystems AG, Aachen, Germany) microaxial flow left and right ventricular assist devices (LVAD/RVAD) were evaluated as they provide circulatory support in the setting of postcardiotomy heart failure refractory to high-dose inotropic and intra-aortic balloon pump (IABP) support.
Methods: Between May 2002 and November 2002, the Recover LVAD was implanted in six patients (64 +/- 11 years) with acute left heart failure following coronary artery bypass procedures. Preoperative left ventricular (LV) ejection fraction was compromised (28% +/- 12%, 12% to 45%). Three patients presented with unstable circulation or cardiogenic shock following acute myocardial infarction, with a predicted mortality rate of 44% +/- 11% (EuroSCORE). Intraoperatively, severe heart failure was present with a more than 70% mortality rate predicted by the IABP score. The Recover RVAD and LVAD were combined to provide biventricular assist device (BVAD) support in one case of post-transplant graft failure.
Results: The Recover LVAD delivered blood flows of up to 5 L/min. A moderate degree of hemolysis and a reduction in platelet count were noted. Four patients were weaned from the LVAD after 169 +/- 34 hours, two of whom remain long-term survivors. Full recovery of graft function allowed weaning of the patient from BVAD support after six days.
Conclusions: The initial experience with the Impella Recover VADs proved the new systems to be advantageous regarding the ease of implantation and device removal, low anticoagulation requirements, and advanced weaning features. In cases of severe heart failure, survival was improved by using LVADs when compared to that predicted by solely continuing IABP and drug support.
Similar articles
-
The Impella Recover microaxial left ventricular assist device reduces mortality for postcardiotomy failure: a three-center experience.J Thorac Cardiovasc Surg. 2004 Mar;127(3):812-22. doi: 10.1016/j.jtcvs.2003.09.055. J Thorac Cardiovasc Surg. 2004. PMID: 15001911
-
Impella recover 100 microaxial left ventricular assist device: the Niguarda experience.Transplant Proc. 2004 Apr;36(3):623-6. doi: 10.1016/j.transproceed.2004.02.051. Transplant Proc. 2004. PMID: 15110614
-
Permanent mechanical circulatory support in patients of advanced age.Eur J Cardiothorac Surg. 2004 Apr;25(4):610-8. doi: 10.1016/j.ejcts.2004.01.017. Eur J Cardiothorac Surg. 2004. PMID: 15037280
-
Percutaneous left ventricular assist devices vs. intra-aortic balloon pump counterpulsation for treatment of cardiogenic shock: a meta-analysis of controlled trials.Eur Heart J. 2009 Sep;30(17):2102-8. doi: 10.1093/eurheartj/ehp292. Epub 2009 Jul 18. Eur Heart J. 2009. PMID: 19617601 Review.
-
Percutaneous left ventricular support devices.Cardiol Clin. 2006 May;24(2):265-75, vii. doi: 10.1016/j.ccl.2006.01.004. Cardiol Clin. 2006. PMID: 16781943 Review.
Cited by
-
Results of extracorporeal life support implementation in routine clinical practice: single center experience.Croat Med J. 2014 Dec;55(6):600-8. doi: 10.3325/cmj.2014.55.600. Croat Med J. 2014. PMID: 25559831 Free PMC article.
-
Contemporary mechanical circulatory support therapy for postcardiotomy shock.Gen Thorac Cardiovasc Surg. 2016 Apr;64(4):183-91. doi: 10.1007/s11748-016-0625-4. Epub 2016 Feb 13. Gen Thorac Cardiovasc Surg. 2016. PMID: 26874519 Review.
-
Structured review of post-cardiotomy extracorporeal membrane oxygenation: part 1-Adult patients.J Heart Lung Transplant. 2019 Nov;38(11):1125-1143. doi: 10.1016/j.healun.2019.08.014. Epub 2019 Aug 10. J Heart Lung Transplant. 2019. PMID: 31522913 Free PMC article. Review.
-
Protected percutaneous coronary intervention with Impella CP in a patient with left main disease, severe left ventricular systolic dysfunction and established hemolysis.Cardiovasc Diagn Ther. 2019 Apr;9(2):194-199. doi: 10.21037/cdt.2019.02.06. Cardiovasc Diagn Ther. 2019. PMID: 31143642 Free PMC article.
-
Adequate systemic perfusion maintained by a CentriMag during acute heart failure.Tex Heart Inst J. 2008;35(3):334-9. Tex Heart Inst J. 2008. PMID: 18941648 Free PMC article.
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Medical