Consciousness monitoring in ventilated patients: bispectral EEG monitors arousal not delirium
- PMID: 15127189
- DOI: 10.1007/s00134-004-2298-1
Consciousness monitoring in ventilated patients: bispectral EEG monitors arousal not delirium
Abstract
Objective: Bispectral index (BIS) is being evaluated as a monitor of consciousness, yet it is unclear what components of consciousness (i.e., arousal vs. content of consciousness) the BIS measures. This study compared BIS levels to well-validated clinical measures of arousal and the presence or absence of delirium.
Design: A prospective, blinded, observational cohort study.
Patients: 124 mechanically ventilated, adult, medical ICU patients.
Measurements and results: Using BIS 3.4 and BIS-XP 4.0 algorithms, BIS values were calculated immediately prior to clinical assessments. The clinical assessments included the Richmond Agitation-Sedation Scale (RASS) and presence or absence of delirium using the Confusion Assessment Method for the ICU. A total of 484 assessments were collected among 124 patients. BIS-XP values demonstrated greater correlation with RASS than BIS 3.4 ( R(2)=0.36 vs. 0.20), although considerable overlap of BIS-XP scores remained across RASS levels. Median BIS-XP values for delirious and nondelirious observations were 74 and 96, respectively, while BIS 3.4 values were 91 and 96, respectively. However, neither BIS 3.4 nor BIS-XP were significantly associated with delirium after controlling for RASS value.
Conclusions: In comparison with clinical measures of arousal in mechanically ventilated patients, BIS-XP algorithm demonstrated stronger correlation with RASS levels than did BIS 3.4, yet marked overlap across different levels of arousal persist using both algorithms. After controlling for level of arousal, neither BIS-XP nor BIS 3.4 algorithms distinguished between the presence and absence of delirium.
Comment in
-
Is disturbance of consciousness an important feature of ICU delirium?Intensive Care Med. 2005 Jun;31(6):887. doi: 10.1007/s00134-005-2642-0. Epub 2005 Apr 21. Intensive Care Med. 2005. PMID: 15843916 No abstract available.
Similar articles
-
Comparison of two bispectral index algorithms in monitoring sedation in postoperative intensive care patients.Crit Care Med. 2005 Mar;33(3):580-4. doi: 10.1097/01.ccm.0000156291.04287.7f. Crit Care Med. 2005. PMID: 15753750 Clinical Trial.
-
Bispectral Index monitoring correlates with sedation scales in brain-injured patients.Crit Care Med. 2004 Dec;32(12):2403-6. doi: 10.1097/01.ccm.0000147442.14921.a5. Crit Care Med. 2004. PMID: 15599143 Clinical Trial.
-
Monitoring sedation status over time in ICU patients: reliability and validity of the Richmond Agitation-Sedation Scale (RASS).JAMA. 2003 Jun 11;289(22):2983-91. doi: 10.1001/jama.289.22.2983. JAMA. 2003. PMID: 12799407
-
Bispectral index monitoring in the intensive care unit provides more signal than noise.Pharmacotherapy. 2005 May;25(5 Pt 2):19S-27S. doi: 10.1592/phco.2005.25.5_part_2.19s. Pharmacotherapy. 2005. PMID: 15899745 Review.
-
Bispectral index as a guide for titration of propofol during procedural sedation among children.Pediatrics. 2005 Jun;115(6):1666-74. doi: 10.1542/peds.2004-1979. Pediatrics. 2005. PMID: 15930231 Review.
Cited by
-
Electroencephalogram-Based Evaluation of Impaired Sedation in Patients with Moderate to Severe COVID-19 ARDS.J Clin Med. 2022 Jun 17;11(12):3494. doi: 10.3390/jcm11123494. J Clin Med. 2022. PMID: 35743572 Free PMC article.
-
Diagnostic considerations regarding pediatric delirium: a review and a proposal for an algorithm for pediatric intensive care units.Intensive Care Med. 2009 Nov;35(11):1843-9. doi: 10.1007/s00134-009-1652-8. Intensive Care Med. 2009. PMID: 19771408 Free PMC article. Review.
-
Year in review in intensive care medicine, 2004. II. Brain injury, hemodynamic monitoring and treatment, pulmonary embolism, gastrointestinal tract, and renal failure.Intensive Care Med. 2005 Feb;31(2):177-88. doi: 10.1007/s00134-004-2552-6. Epub 2005 Jan 28. Intensive Care Med. 2005. PMID: 15678311 Review. No abstract available.
-
Systematic review and meta-analysis of the correlation between bispectral index (BIS) and clinical sedation scales: Toward defining the role of BIS in critically ill patients.Pharmacotherapy. 2022 Aug;42(8):667-676. doi: 10.1002/phar.2712. Epub 2022 Jun 30. Pharmacotherapy. 2022. PMID: 35707961 Free PMC article.
-
The Presence/Absence of an Awake-State Dominant EEG Rhythm in Delirious Patients Is Related to Different Symptoms of Delirium Evaluated by the Intensive Care Delirium Screening Checklist (ICDSC).Sensors (Basel). 2024 Dec 19;24(24):8097. doi: 10.3390/s24248097. Sensors (Basel). 2024. PMID: 39771830 Free PMC article.