Monitoring sub-nanogram amount of acetylspiramycin in human urine using flow injection analysis with chemiluminescence detection
- PMID: 15127585
Monitoring sub-nanogram amount of acetylspiramycin in human urine using flow injection analysis with chemiluminescence detection
Abstract
Aim: To establish a new and simple flow injection method for the rapid determination of acetylspiramycin (ASPM).
Methods: ASPM was determined by chemiluminescence (CL) method combined with flow injection (FI) technology, which was based on the inhibitive effect of ASPM on the chemiluminescence reaction of the luminol-K3Fe (CN)6 system.
Results: The decrease of chemiluminescence intensity was proportional to the logarithm of ASPM concentration (0.1-100) microgram.L-1, the detection limit was 40 ng.L-1 (3 sigma). The whole process, including sampling and washing, could be completed in 0.5 min with a RSD less than 3.0% (n = 5).
Conclusion: The FI-CL method is of both high sensitivity and good selectivity giving a throughput of 120 h-1. The proposed method was applied successfully to the determination of ASPM in pharmaceutical preparations and human urine without any pre-treatment. It was found that the ASPM concentration reached its maximum after being orally administrated for two hours.
Similar articles
-
Ultrasensitive assay of clindamycin in medicine and bio-fluids with chemiluminescence detection.Luminescence. 2007 Nov-Dec;22(6):534-9. doi: 10.1002/bio.996. Luminescence. 2007. PMID: 17768707
-
Chemiluminescence flow-through sensor for the determination of vitamin B2 using controlled-reagent-release technology.Yao Xue Xue Bao. 2002 Oct;37(10):793-7. Yao Xue Xue Bao. 2002. PMID: 12567864
-
Rapid determination of clindamycin in medicine with myoglobin-luminol chemiluminescence system.J Pharm Biomed Anal. 2006 May 3;41(2):667-70. doi: 10.1016/j.jpba.2005.12.027. Epub 2006 Jan 24. J Pharm Biomed Anal. 2006. PMID: 16439092
-
[Recent progress in flow-injection chemiluminescence for pharmaceutical analysis].Guang Pu Xue Yu Guang Pu Fen Xi. 2006 May;26(5):816-20. Guang Pu Xue Yu Guang Pu Fen Xi. 2006. PMID: 16883844 Review. Chinese.
-
Flow injection assays with chemiluminescence and bioluminescence detection--a review.J Biolumin Chemilumin. 1993 Jul-Aug;8(4):183-99. doi: 10.1002/bio.1170080402. J Biolumin Chemilumin. 1993. PMID: 8372703 Review.
Publication types
MeSH terms
Substances
LinkOut - more resources
Medical