Skip to main page content
U.S. flag

An official website of the United States government

Dot gov

The .gov means it’s official.
Federal government websites often end in .gov or .mil. Before sharing sensitive information, make sure you’re on a federal government site.

Https

The site is secure.
The https:// ensures that you are connecting to the official website and that any information you provide is encrypted and transmitted securely.

Access keys NCBI Homepage MyNCBI Homepage Main Content Main Navigation
. 2004 May;45(5):797-801.

Impact of patient weight and emission scan duration on PET/CT image quality and lesion detectability

Affiliations
  • PMID: 15136629
Free article

Impact of patient weight and emission scan duration on PET/CT image quality and lesion detectability

Benjamin S Halpern et al. J Nucl Med. 2004 May.
Free article

Abstract

This study was performed to prospectively evaluate fast PET/CT imaging protocols using lutetium oxyorthosilicate (LSO) detector technology and 3-dimensional (3D) image-acquisition protocols.

Methods: Fifty-seven consecutive patients (30 male, 27 female; mean age, 58.6 +/- 15.7 y) were enrolled in the study. After intravenous injection of 7.77 MBq (0.21 mCi) of (18)F-FDG per kilogram, a standard whole-body CT study (80-110 s) and PET emission scan were acquired for 4 min/bed position in 49 patients and 3 min/bed position in 8 patients. One-minute-per-bed-position data were then extracted from the 3- or 4-min/bed position scans to reconstruct single-minute/bed position scans for each patient. Patients were subgrouped according to weight as follows: <59 kg (<130 lb; n = 15), 59-81 kg (130-179 lb; n = 33), and >or=82 kg (>or=180 lb; n = 9). Three experienced observers recorded numbers and locations of lesion by consensus and independently rated image quality as good, moderate, poor, or nondiagnostic.

Results: The observers analyzed 220 reconstructed whole-body PET images from 57 patients. They identified 114 lesions ranging in size from 0.7 to 7.0 cm on the 3- (n = 8) and 4-min/bed position images (n = 49). Of these, only 4 were missed on the 1-min/bed position scans, and all lesions were identified on the corresponding 2-min/bed position images. One- and 2-min/bed position image quality differed significantly from the 4-min/bed position image reference (P < 0.05).

Conclusion: LSO PET detector technology permits fast 3D imaging protocols whereby weight-based emission scan durations ranging from 1 to 3 min/bed position provide similar lesion detectability when compared with 4-min/bed position images.