The biological reaction to polyethylene wear debris can be related with oxidation of the UHMWPE cups
Abstract
Background: Oxidised UHMWPE due to gamma irradiation in air has a greater susceptibility to wear than non-oxidised UHMWPE (ethylene oxide, EtO).
Aim: To evaluate, the biological reaction of loose implants with oxidised and non-oxidised PE components.
Materials and methods: Ten loose PE cups sterilised by EtO (group 1) and 13 sterilised by gamma irradiation in air (group 2) were studied. PE wear and oxidation were related to biological reaction.
Results: Group 1: PE cups had low wear score and no oxidation. In the oldest implants, a few PE particles and macrophages were observed in the interface membrane. Group 2: PE inserts had medium-to-high PE wear score and variable oxidation. In the interface membrane, the number and total area of PE particles were high and variable; giant cells were more numerous than macrophages.
Conclusion: Junctional tissues around loose oxidised PE components contain more PE debris and giant cells than membranes around PE components that are not oxidised.
Similar articles
-
Very low wear of non-remelted highly cross-linked polyethylene cups: an RSA study lasting up to 6 years.Acta Orthop. 2007 Dec;78(6):739-45. doi: 10.1080/17453670710014509. Acta Orthop. 2007. PMID: 18236179
-
Comparative wear and wear debris under three different counterface conditions of crosslinked and non-crosslinked ultra high molecular weight polyethylene.Biomed Mater Eng. 2001;11(1):23-35. Biomed Mater Eng. 2001. PMID: 11281576
-
Wear behaviour of cross-linked polyethylene assessed in vitro under severe conditions.Biomaterials. 2005 Jun;26(16):3259-67. doi: 10.1016/j.biomaterials.2004.07.070. Biomaterials. 2005. PMID: 15603821
-
New polymer materials in total hip arthroplasty. Evaluation with radiostereometry, bone densitometry, radiography and clinical parameters.Acta Orthop Suppl. 2005 Feb;76(315):3-82. Acta Orthop Suppl. 2005. PMID: 15790289 Review.
-
Implant wear mechanisms--basic approach.Biomed Mater. 2008 Dec;3(4):042001. doi: 10.1088/1748-6041/3/4/042001. Epub 2008 Sep 25. Biomed Mater. 2008. PMID: 18824778 Review.
Cited by
-
Vitamin E-stabilized UHMWPE for total joint implants: a review.Clin Orthop Relat Res. 2011 Aug;469(8):2286-93. doi: 10.1007/s11999-010-1717-6. Clin Orthop Relat Res. 2011. PMID: 21132413 Free PMC article. Review.
-
Ultra-high molecular weight polyethylene (UHMWPE) for hip and knee arthroplasty: The present and the future.J Orthop. 2021 Apr 23;25:98-106. doi: 10.1016/j.jor.2021.04.004. eCollection 2021 May-Jun. J Orthop. 2021. PMID: 33994706 Free PMC article. Review.
-
UHMWPE for arthroplasty: past or future?J Orthop Traumatol. 2009 Mar;10(1):1-8. doi: 10.1007/s10195-008-0038-y. Epub 2008 Dec 24. J Orthop Traumatol. 2009. PMID: 19384628 Free PMC article.
-
Mediators of the inflammatory response to joint replacement devices.Nat Rev Rheumatol. 2011 Sep 6;7(10):600-8. doi: 10.1038/nrrheum.2011.128. Nat Rev Rheumatol. 2011. PMID: 21894210 Review.
-
Sterilization of Polymeric Implants: Challenges and Opportunities.ACS Appl Bio Mater. 2022 Nov 21;5(11):5077-5088. doi: 10.1021/acsabm.2c00793. Epub 2022 Nov 1. ACS Appl Bio Mater. 2022. PMID: 36318175 Free PMC article. Review.