Designing optically tracked instruments for image-guided surgery
- PMID: 15147007
- DOI: 10.1109/tmi.2004.825614
Designing optically tracked instruments for image-guided surgery
Abstract
Most image-guided surgery (IGS) systems track the positions of surgical instruments in the physical space occupied by the patient. This task is commonly performed using an optical tracking system that determines the positions of fiducial markers such as infrared-emitting diodes or retroreflective spheres that are attached to the instrument. Instrument tracking error is an important component of the overall IGS system error. This paper is concerned with the effect of fiducial marker configuration (number and spatial distribution) on tip position tracking error. Statistically expected tip position tracking error is calculated by applying results from the point-based registration error theory developed by Fitzpatrick et al. Tracking error depends not only on the error in localizing the fiducials, which is the error value generally provided by manufacturers of optical tracking systems, but also on the number and spatial distribution of the tracking fiducials and the position of the instrument tip relative to the fiducials. The theory is extended in two ways. First, a formula is derived for the special case in which the fiducials and the tip are collinear. Second, the theory is extended for the case in which there is a composition of transformations, as is the situation for tracking an instrument relative to a coordinate reference frame (i.e., a set of fiducials attached to the patient). The derivation reveals that the previous theory may be applied independently to the two transformations; the resulting independent components of tracking error add in quadrature to give the overall tracking error. The theoretical results are verified with numerical simulations and experimental measurements. The results in this paper may be useful for the design of optically tracked instruments for image-guided surgery; this is illustrated with several examples.
Similar articles
-
Implementation, calibration and accuracy testing of an image-enhanced endoscopy system.IEEE Trans Med Imaging. 2002 Dec;21(12):1524-35. doi: 10.1109/tmi.2002.806597. IEEE Trans Med Imaging. 2002. PMID: 12588036
-
Passive markers for ultrasound tracking of surgical instruments.Med Image Comput Comput Assist Interv. 2005;8(Pt 2):41-8. doi: 10.1007/11566489_6. Med Image Comput Comput Assist Interv. 2005. PMID: 16685941
-
Determining the position of a patient reference from C-Arm views for image guided navigation.Int J Comput Assist Radiol Surg. 2011 Mar;6(2):217-27. doi: 10.1007/s11548-010-0502-z. Epub 2010 Jul 18. Int J Comput Assist Radiol Surg. 2011. PMID: 20640603
-
Tracking of the liver for navigation in open surgery.Int J Comput Assist Radiol Surg. 2010 May;5(3):229-35. doi: 10.1007/s11548-009-0395-x. Epub 2009 Aug 7. Int J Comput Assist Radiol Surg. 2010. PMID: 20033499 Review.
-
Optically guided patient positioning techniques.Semin Radiat Oncol. 2005 Jul;15(3):192-201. doi: 10.1016/j.semradonc.2005.01.004. Semin Radiat Oncol. 2005. PMID: 15983944 Review.
Cited by
-
Advances of surgical robotics: image-guided classification and application.Natl Sci Rev. 2024 Jun 6;11(9):nwae186. doi: 10.1093/nsr/nwae186. eCollection 2024 Sep. Natl Sci Rev. 2024. PMID: 39144738 Free PMC article. Review.
-
General approach to first-order error prediction in rigid point registration.IEEE Trans Med Imaging. 2011 Mar;30(3):679-93. doi: 10.1109/TMI.2010.2091513. Epub 2010 Nov 11. IEEE Trans Med Imaging. 2011. PMID: 21075718 Free PMC article.
-
Design and validation of an open-source library of dynamic reference frames for research and education in optical tracking.J Med Imaging (Bellingham). 2018 Apr;5(2):021215. doi: 10.1117/1.JMI.5.2.021215. Epub 2018 Feb 9. J Med Imaging (Bellingham). 2018. PMID: 29487887 Free PMC article.
-
A random phased-array for MR-guided transcranial ultrasound neuromodulation in non-human primates.Phys Med Biol. 2018 May 17;63(10):105016. doi: 10.1088/1361-6560/aabeff. Phys Med Biol. 2018. PMID: 29667598 Free PMC article.
-
Magnetic tracking in the operation room using the da Vinci(®) telemanipulator is feasible.J Robot Surg. 2013 Mar;7(1):59-64. doi: 10.1007/s11701-012-0347-2. Epub 2012 Mar 24. J Robot Surg. 2013. PMID: 23440620 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources