Abstract
Introduction: Shot peening and grit blasting techniques are used in several surface modifications for producing hip endoprostheses. Corund blasting techniques using highly pure Al2O3 particles to create a rough surface of titanium implants in cementless total hip arthroplasty achieve better osteointegration. An increasing number of publications in maxillofacial surgery and orthopedic surgery show that there is a significant contamination on corund-blasted surfaces. The latest research studies reported an effect of contaminant particles on early failure of endoprostheses associated with third body wear. The aim of our work was to evaluate the amount and particle size of Al2O3 on the implant surface.
Material and method: The surface of four different geometries (five parts each, ARR titanium acetabular reinforcement ring, anisotropic Vektor cup, Vektor titanium stem and modular MRP Titan stem) were analyzed with respect to Al2O3 particles. Grit blasting was performed with Al2O3 particles, Alodur SWSK, Fa. Treibacher. Field emission scanning electron microscopy (LEO 1525) was used for the detection of the Al2O3 particles at the implant surface with a backscattered electron detector. The particle size distribution as well as the average area of the covered surface was calculated with image analyzing software (analySIS, Soft Imaging System GmbH).
Results: The surface of the anisotropic Vektor cup was contaminated at an average of 41.7+/-4.9%, the Vektor titanium stem at an average of 33.3+/-4.7%, the (MRP) Titan stem at an average of 30.6+/-4.2%, and the ARR titanium acetabular reinforcement ring at an average of 23.2+/-1.6 with Al2O3 particles over the whole surface orientated to the bone.
Discussion: The results of this study clearly show that there is contamination of rough titanium surfaces with Al2O3 particles at an unexpectedly high percentage. With respect to third body wear in total hip arthroplasty, further studies are necessary to minimize contamination of roughened surfaces and maintain sufficient roughness for osteointegration.
Similar articles
-
[The use of a thermomechanical cleaning procedure for removal of residual particles of corund blasted or glass bead peened implants in total hip arthoplasty].Zentralbl Chir. 2005 Aug;130(4):346-52. doi: 10.1055/s-2005-836801. Zentralbl Chir. 2005. PMID: 16103960 German.
-
[Surface characterisation of shot peened implants with glas beads in total hip arthroplasty].Zentralbl Chir. 2004 Jun;129(3):225-9. doi: 10.1055/s-2004-822740. Zentralbl Chir. 2004. PMID: 15237332 German.
-
[Comparative surface examinations of morse taper junctions of the MRP-Titan stem shot peened with glass beads].Biomed Tech (Berl). 2004 Dec;49(12):334-9. doi: 10.1515/BMT.2004.062. Biomed Tech (Berl). 2004. PMID: 15655925 German.
-
[Residual particle free rough surfaces after rough blasting with steel grit in total hip arthroplasty].Biomed Tech (Berl). 2005 Dec;50(12):404-7. doi: 10.1515/BMT.2005.056. Biomed Tech (Berl). 2005. PMID: 16429944 German.
-
[The role of surface roughness in promoting osteointegration].Refuat Hapeh Vehashinayim (1993). 2003 Jul;20(3):8-19, 98. Refuat Hapeh Vehashinayim (1993). 2003. PMID: 14515625 Review. Hebrew.
Cited by
-
High Revision Rates of a Cementless Beta-Titanium Alloy Stem with Contamination-Free Roughened Surface in Primary Total Hip Arthroplasty.J Clin Med. 2020 Jul 7;9(7):2138. doi: 10.3390/jcm9072138. J Clin Med. 2020. PMID: 32645917 Free PMC article.
-
Unexpected early loosening of rectangular straight femoral Zweymüller stems with an alumina-reduced surface after total hip arthroplasty-a prospective, double-blind, randomized controlled trial.J Orthop Traumatol. 2024 Mar 2;25(1):12. doi: 10.1186/s10195-023-00743-1. J Orthop Traumatol. 2024. PMID: 38430413 Free PMC article. Clinical Trial.
-
A cementless hip system with a new surface for osseous integration.Int Orthop. 2014 Apr;38(4):703-9. doi: 10.1007/s00264-013-2135-7. Epub 2013 Oct 11. Int Orthop. 2014. PMID: 24114250 Free PMC article.
-
Biomaterials in orthopaedics.J R Soc Interface. 2008 Oct 6;5(27):1137-58. doi: 10.1098/rsif.2008.0151. J R Soc Interface. 2008. PMID: 18667387 Free PMC article. Review.
-
Surface modification of titanium implants using bioactive glasses with air abrasion technologies.J Mater Sci Mater Med. 2007 Dec;18(12):2291-6. doi: 10.1007/s10856-007-3137-z. Epub 2007 Jun 12. J Mater Sci Mater Med. 2007. PMID: 17562133
References
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Miscellaneous