Improved hemodynamic performance with a novel chest compression device during treatment of in-hospital cardiac arrest
- PMID: 15172705
- DOI: 10.1016/j.resuscitation.2004.01.025
Improved hemodynamic performance with a novel chest compression device during treatment of in-hospital cardiac arrest
Abstract
Introduction: The purpose of this pilot clinical study was to determine if a novel chest compression device would improve hemodynamics when compared to manual chest compression during cardiopulmonary resuscitation (CPR) in humans. The device is an automated self-adjusting electromechanical chest compressor based on AutoPulse technology (Revivant Corporation) that uses a load distributing compression band (A-CPR) to compress the anterior chest.
Methods: A total of 31 sequential subjects with in-hospital sudden cardiac arrest were screened with institutional review board approval. All subjects had received prior treatment for cardiac disease and most had co-morbidities. Subjects were included following 10 min of failed standard advanced life support (ALS) protocol. Fluid-filled catheters were advanced into the thoracic aorta and the right atrium and placement was confirmed by pressure waveforms and chest radiograph. The coronary perfusion pressure (CPP) was measured as the difference between the aortic and right atrial pressure during the chest compression's decompressed state. Following 10 min of failed ALS and catheter placement, subjects received alternating manual and A-CPR chest compressions for 90 s each. Chest compressions were administered without ventilation pauses at 100 compressions/min for manual CPR and 60 compressions/min for A-CPR. All subjects were intubated and ventilated by bag-valve at 12 breaths/min between compressions. Epinephrine (adrenaline) (1mg i.v. bolus) was given at the request of the attending physician at 3-5 min intervals. Usable pressure signals were present in 16 patients (68 +/- 6 years, 5 female), and data are reported from those patients only. A-CPR chest compressions increased peak aortic pressure when compared to manual chest compression (153 +/- 28 mmHg versus 115 +/- 42 mmHg, P < 0.0001, mean +/- S.D.). Similarly, A-CPR increased peak right atrial pressure when compared to manual chest compression (129 +/- 32 mmHg versus 83 +/- 40 mmHg, P < 0.0001). Furthermore, A-CPR increased CPP over manual chest compression (20 +/- 12 mmHg versus 15 +/- 11 mmHg, P < 0.015). Manual chest compressions were of consistent high quality (51 +/- 20 kg) and in all cases met or exceeded American Heart Association guidelines for depth of compression.
Conclusion: Previous research has shown that increased CPP is correlated to increased coronary blood flow and increased rates of restored native circulation from sudden cardiac arrest. The A-CPR system using AutoPulse technology demonstrated increased coronary perfusion pressure over manual chest compression during CPR in this terminally ill patient population.
Similar articles
-
A new device producing manual sternal compression with thoracic constraint for cardiopulmonary resuscitation.Resuscitation. 2006 May;69(2):295-301. doi: 10.1016/j.resuscitation.2005.07.025. Epub 2006 Feb 2. Resuscitation. 2006. PMID: 16457933
-
Cardiopulmonary resuscitation with a novel chest compression device in a porcine model of cardiac arrest: improved hemodynamics and mechanisms.J Am Coll Cardiol. 2004 Dec 7;44(11):2214-20. doi: 10.1016/j.jacc.2004.08.061. J Am Coll Cardiol. 2004. PMID: 15582320
-
Incomplete chest wall decompression: a clinical evaluation of CPR performance by EMS personnel and assessment of alternative manual chest compression-decompression techniques.Resuscitation. 2005 Mar;64(3):353-62. doi: 10.1016/j.resuscitation.2004.10.007. Resuscitation. 2005. PMID: 15733766 Clinical Trial.
-
Could we use abdominal compressions rather than chest compression in patients who arrest after cardiac surgery?Interact Cardiovasc Thorac Surg. 2009 Jan;8(1):148-51. doi: 10.1510/icvts.2008.195974. Epub 2008 Oct 23. Interact Cardiovasc Thorac Surg. 2009. PMID: 18948306 Review.
-
Hemodynamics of cardiac massage.Emerg Med Clin North Am. 1983 Dec;1(3):501-13. Emerg Med Clin North Am. 1983. PMID: 6396069 Review.
Cited by
-
Grade IV Liver Injury Following Mechanical Cardiopulmonary Resuscitation with Postoperative Three-dimensional Evaluation.J Emerg Trauma Shock. 2020 Oct-Dec;13(4):306-308. doi: 10.4103/JETS.JETS_28_20. Epub 2020 Dec 7. J Emerg Trauma Shock. 2020. PMID: 33897149 Free PMC article.
-
Continuous mechanical chest compression-assisted percutaneous coronary intervention in a patient with cardiac arrest complicating acute myocardial infarction.Chin Med J (Engl). 2015 Mar 20;128(6):846-8. doi: 10.4103/0366-6999.152692. Chin Med J (Engl). 2015. PMID: 25758285 Free PMC article. No abstract available.
-
Retrospective Study Using Computed Tomography to Compare Sufficient Chest Compression Depth for Cardiopulmonary Resuscitation in Obese Patients.J Am Heart Assoc. 2019 Dec 3;8(23):e013948. doi: 10.1161/JAHA.119.013948. Epub 2019 Nov 26. J Am Heart Assoc. 2019. PMID: 31766971 Free PMC article.
-
Blood flow maintenance by cardiac massage during cardiopulmonary resuscitation: Classical theories, newer hypotheses, and clinical utility of mechanical devices.J Intensive Care Soc. 2019 Feb;20(1):2-10. doi: 10.1177/1751143718778486. Epub 2018 Jun 18. J Intensive Care Soc. 2019. PMID: 30792756 Free PMC article. Review.
-
Monitoring of cerebral oxygen saturation during resuscitation in out-of-hospital cardiac arrest: a feasibility study in a physician staffed emergency medical system.Scand J Trauma Resusc Emerg Med. 2014 Oct 5;22:58. doi: 10.1186/s13049-014-0058-y. Scand J Trauma Resusc Emerg Med. 2014. PMID: 25286829 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Medical
Miscellaneous