Impact of photon energy recovery on the assessment of left ventricular volume using myocardial perfusion SPECT
- PMID: 15173778
- DOI: 10.1016/j.nuclcard.2004.03.001
Impact of photon energy recovery on the assessment of left ventricular volume using myocardial perfusion SPECT
Abstract
Background: Photon energy recovery (PER) is a spectral deconvolution technique validated for scatter removal in patients and phantom studies. The purpose of this study was to examine the impact of PER on left ventricular volume measurement based on myocardial perfusion single photon emission computed tomography (SPECT).
Methods and results: SPECT acquisitions were performed by use of a static cardiac phantom and in 25 patients after a rest injection of technetium 99m sestamibi by use of multiple energy windows (126-136, 137-144, and 145-154 keV). Data were successively reconstructed with and without PER, by use of iterative reconstruction and post-processing filtering (Butterworth filter; order, 5; cutoff, 0.30 cycles/pixel). Image contrast was evaluated in reconstructed data, and volumes were calculated by use of QGS. PER increased reconstructed image contrast from 62% +/- 2.7% to 84.3% +/- 5.7% in the phantom studies (P <.0001) and from 49% +/- 2% to 73% +/- 2% in patients (P <.0001). Although it remained underestimated (P <.0001), phantom volume was higher after PER correction compared with uncorrected data (50.9 +/- 0.8 mL vs 44.6 +/- 1 mL, P <.0001). The error in volume measurement was decreased by PER correction (16.6% +/- 1.3% vs 27% +/- 1.7% [uncorrected data], P <.0001). In patients, left ventricular volume increased from 83 +/- 10 mL to 91 +/- 10 mL (P <.0001), and the PER-induced volume increase was correlated with the image contrast increase (r = 0.61, P =.001). Finally, the percentage of volume increase was higher in patients with small left ventricular volumes.
Conclusions: PER has a significant impact on image contrast and left ventricular volume measurement by use of perfusion SPECT. PER improves the accuracy of phantom volume assessment. In patients, volume increase is correlated to image contrast increase and is higher in those with small ventricles.
Similar articles
-
Impact of Wiener filter in determining the left ventricular volume and ejection fraction using thallium-201 gated SPECT.Nucl Med Commun. 2003 Aug;24(8):907-14. doi: 10.1097/01.mnm.0000084587.29433.ca. Nucl Med Commun. 2003. PMID: 12869824 Clinical Trial.
-
Comparison of two software in gated myocardial perfusion single photon emission tomography, for the measurement of left ventricular volumes and ejection fraction, in patients with and without perfusion defects.Hell J Nucl Med. 2007 Jan-Apr;10(1):19-23. Hell J Nucl Med. 2007. PMID: 17450245
-
Validation of 4D-MSPECT and QGS for quantification of left ventricular volumes and ejection fraction from gated 99mTc-MIBI SPET: comparison with cardiac magnetic resonance imaging.Eur J Nucl Med Mol Imaging. 2004 Apr;31(4):482-90. doi: 10.1007/s00259-003-1411-y. Epub 2004 Jan 14. Eur J Nucl Med Mol Imaging. 2004. PMID: 14722679 Clinical Trial.
-
Tracking a therapeutic response: how reliable are serial measurements of LV perfusion and function?J Nucl Cardiol. 2012 Apr;19(2):360-3. doi: 10.1007/s12350-012-9518-9. J Nucl Cardiol. 2012. PMID: 22297852 Free PMC article. Review. No abstract available.
-
Advances in technical aspects of myocardial perfusion SPECT imaging.J Nucl Cardiol. 2009 Mar-Apr;16(2):255-76. doi: 10.1007/s12350-009-9052-6. Epub 2009 Feb 26. J Nucl Cardiol. 2009. PMID: 19242769 Review. No abstract available.
Cited by
-
Segmental and global left ventricular function assessment using gated SPECT with a semiconductor Cadmium Zinc Telluride (CZT) camera: phantom study and clinical validation vs cardiac magnetic resonance.J Nucl Cardiol. 2014 Aug;21(4):712-22. doi: 10.1007/s12350-014-9899-z. Epub 2014 May 9. J Nucl Cardiol. 2014. PMID: 24810429
-
Transplant allograft vasculopathy: Role of multimodality imaging in surveillance and diagnosis.J Nucl Cardiol. 2016 Aug;23(4):713-27. doi: 10.1007/s12350-015-0373-3. Epub 2015 Dec 28. J Nucl Cardiol. 2016. PMID: 26711101 Review.
References
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Medical