Comparison of ultrafiltration and solid phase extraction for the separation of free and protein-bound serum copper for the Wilson's disease diagnosis
- PMID: 15193985
- DOI: 10.1016/j.cccn.2004.03.001
Comparison of ultrafiltration and solid phase extraction for the separation of free and protein-bound serum copper for the Wilson's disease diagnosis
Abstract
Background: The determination of the ratio free/protein-bound serum copper along with urinary copper can be used as a preliminary test for the Wilson's Disease diagnosis. In this work, the determination of these copper fractions in serum samples was carried out in two different ways; after separation of the copper bound to proteins from the free fraction by a column for protein adsorption and by ultrafiltration. As proteins can be adsorbed onto plastic polymeric surfaces, polyethylene (PE) with different molecular weights in powder form was investigated for protein adsorption. A small column was adapted in a flow system to carry out a solid-phase extraction (SPE) on-line. Preliminary experiments defined conditions for protein retention and elution and column saturation. Good performance was achieved using Mg(NO3)2 solution as carrier and methanol as eluent. The presence of proteins in both fraction (column effluent and eluate) was checked by the Coomassie Brilliant Blue test. Copper was measured by graphite furnace atomic absorption spectrometry. The measurement in the column effluent furnished the free-fraction of copper while the copper measured in the eluate the bound-fraction. The method was compared with ultrafiltration (20 kDa), measuring the free-copper in the ultrafiltrate. For the determination of protein-bound copper, the copper found in the ultrafitrate was discounted from the total copper measured in the sample.
Results: Serum samples of 10 individuals were analyzed by both methods with good agreement of the results. The regression plots, obtained by analysing the samples by both methods, presented r2 and slope of 0.97 and 0.96 for free copper and 1.00 and 1.00 for bound copper, respectively. Protein-bound copper (PB) concentrations ranged from 74 to 2074 microg/l and free-copper (F) from 22 to 54 microg/l. The ratio F/PB, calculated from SPE data, was 29.7% for one individual, with Wilson Disease well-characterized, and ranged from 1.2% to 5.2% for the others.
Conclusion: The SPE method performed well in terms of accuracy and precision, and showed good agreement with the UF. Advantages of SPE are small sample volume (50 microl), separation carried out in 10 min, and the use of the same column for several analyses.
Copyright 2004 Elsevier B.V.
Similar articles
-
The determination of free and protein-bound haemoglobin in plasma using a combination of HPLC and absorption spectrometry.Clin Lab. 2001;47(5-6):279-88. Clin Lab. 2001. PMID: 11405606
-
Wilson's disease: the importance of measuring serum caeruloplasmin non-immunologically.Ann Clin Biochem. 2003 Mar;40(Pt 2):115-21. doi: 10.1258/000456303763046021. Ann Clin Biochem. 2003. PMID: 12662398 Review.
-
Direct sample introduction of wines in graphite furnace atomic absorption spectrometry for the simultaneous determination of arsenic, cadmium, copper and lead content.Talanta. 2008 Jul 30;76(3):627-34. doi: 10.1016/j.talanta.2008.04.014. Epub 2008 Apr 16. Talanta. 2008. PMID: 18585331
-
Performance of polymeric materials for solid phase extraction prior to chromatographic analysis.J Chromatogr B Analyt Technol Biomed Life Sci. 2009 Jan 15;877(3):277-84. doi: 10.1016/j.jchromb.2008.12.025. Epub 2008 Dec 24. J Chromatogr B Analyt Technol Biomed Life Sci. 2009. PMID: 19117810
-
New methods of protein purification. Affinity ultrafiltration.Biochemistry (Mosc). 1999 Aug;64(8):849-56. Biochemistry (Mosc). 1999. PMID: 10498798 Review.
Cited by
-
Adjusting copper concentrations for caeruloplasmin levels in routine clinical practice.J Clin Pathol. 2006 Aug;59(8):867-9. doi: 10.1136/jcp.2005.034876. Epub 2006 Apr 27. J Clin Pathol. 2006. PMID: 16644878 Free PMC article.
-
A fluorometric assay to determine labile copper(II) ions in serum.Sci Rep. 2023 Aug 7;13(1):12807. doi: 10.1038/s41598-023-39841-9. Sci Rep. 2023. PMID: 37550465 Free PMC article.
-
Penicillamine increases free copper and enhances oxidative stress in the brain of toxic milk mice.PLoS One. 2012;7(5):e37709. doi: 10.1371/journal.pone.0037709. Epub 2012 May 21. PLoS One. 2012. PMID: 22629446 Free PMC article.
-
A caged imidazopyrazinone for selective bioluminescence detection of labile extracellular copper(ii).Chem Sci. 2022 Mar 23;13(15):4352-4363. doi: 10.1039/d1sc07177g. eCollection 2022 Apr 13. Chem Sci. 2022. PMID: 35509459 Free PMC article.
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical
Research Materials