The impact of pump speed and inlet cannulation site on left ventricular unloading with a rotary blood pump
- PMID: 15209860
- DOI: 10.1111/j.1525-1594.2004.07374.x
The impact of pump speed and inlet cannulation site on left ventricular unloading with a rotary blood pump
Abstract
Background: Ventricular assist devices are gaining ground in the therapeutic treatment of chronic heart failure. These devices are sometimes used as a bridge to recovery by unloading the left ventricle (LV) and restoring its function. It is therefore important to preserve the heart muscle and apply less invasive implantation methods.
Methods: In this study ventricular unloading was achieved in 7 healthy sheep with a rotary blood pump at different pump flow levels. Ventricular cannulation via the left atrium (LA) and through the mitral valve was compared to atrial cannulation. The unloading of the heart was assessed with LV pressure-volume loops, derived energetic parameters, and an estimate of LV wall stress.
Results: No significant difference between the cannulations was found for any flow or pressure. LA cannulation, however, resulted in significantly lower stroke volumes and stroke work for all pump flow levels. Irrespective of cannulation site, LV volumes and energetic parameters showed a significant decrease with increasing pump flow.
Conclusion: LV assist with a rotary blood pump can provide sufficient unloading with atrial cannulation.
Similar articles
-
Autosynchronized systolic unloading during left ventricular assist with a centrifugal pump.J Thorac Cardiovasc Surg. 2003 Feb;125(2):353-60. doi: 10.1067/mtc.2003.100. J Thorac Cardiovasc Surg. 2003. PMID: 12579105
-
Effects of pulsatile- and continuous-flow left ventricular assist devices on left ventricular unloading.J Heart Lung Transplant. 2008 Mar;27(3):261-7. doi: 10.1016/j.healun.2007.12.001. J Heart Lung Transplant. 2008. PMID: 18342746
-
Pulsatile control of rotary blood pumps: Does the modulation waveform matter?J Thorac Cardiovasc Surg. 2012 Oct;144(4):970-7. doi: 10.1016/j.jtcvs.2012.02.015. Epub 2012 Mar 12. J Thorac Cardiovasc Surg. 2012. PMID: 22418246
-
Augmentation of dilated failing left ventricular stroke work by a physiological cardiac assist device.Ann N Y Acad Sci. 2004 May;1015:379-90. doi: 10.1196/annals.1302.033. Ann N Y Acad Sci. 2004. PMID: 15201176 Review.
-
Current axial-flow devices--the HeartMate II and Jarvik 2000 left ventricular assist devices.Semin Thorac Cardiovasc Surg. 2008 Fall;20(3):264-72. doi: 10.1053/j.semtcvs.2008.08.001. Semin Thorac Cardiovasc Surg. 2008. PMID: 19038737 Review.
Cited by
-
The future of adult cardiac assist devices: novel systems and mechanical circulatory support strategies.Cardiol Clin. 2011 Nov;29(4):559-82. doi: 10.1016/j.ccl.2011.08.013. Cardiol Clin. 2011. PMID: 22062206 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources