New blood volume monitoring method for hemodialysis: A-V pressure gradient measurement by synchronized one-point reading
- PMID: 15209863
- DOI: 10.1111/j.1525-1594.2004.07302.x
New blood volume monitoring method for hemodialysis: A-V pressure gradient measurement by synchronized one-point reading
Abstract
During hemodialysis, rapid ultrafiltration often causes symptomatic hypotension. To predict the occurrence of volume-dependent hypotension as early as possible, continuous hematocrit monitoring with the Crit-Line noninvasive monitor has been widely used to measure blood volume changes during hemodialysis. As another potential method of monitoring blood volume variations, we studied blood viscosity, which is theoretically associated with the pressure gradient across the dialyzer. Blood viscosity (calculated by the Hugen-Poiseuille formula) is a major determinant of the blood flow rate and is associated with the pressure difference between the postpump arterial (A) and venous (V) pressures. The A-V pressure gradient fluctuates due to pump pulsation, so we minimized this noise by always reading the pressure gradient at the same point out of 1400 partitions on the rotary pump. To test this synchronized one-point reading method, the A-V pressure gradient was measured using 3 different xanthan gum solutions and was found to be linearly proportional to the model blood flow rate. In an experimental dialysis system using a xanthan gum solution (300 mg/L), the A-V pressure gradient showed a gradual linear increase along with the ultrafiltration rate up to 1 L/h as the viscosity slowly increased in the dialyzer. The changes of blood volume shown by this method were significantly correlated with data obtained using the Crit-Line in 8 patients undergoing hemodialysis. This simple and inexpensive method may allow monitoring of blood volume changes and thus provide data that are beneficial for fluid management in hemodialysis patients suffering from clinical dialysis intolerance.
Similar articles
-
Haemodialysis with on-line monitoring equipment: tools or toys?Nephrol Dial Transplant. 2005 Jan;20(1):22-33. doi: 10.1093/ndt/gfh555. Nephrol Dial Transplant. 2005. PMID: 15632348 Review.
-
Development of a new method for monitoring blood purification: the blood flow analysis of the head and foot by laser Doppler blood flowmeter during hemodialysis.Hemodial Int. 2005 Jan;9(1):56-62. doi: 10.1111/j.1492-7535.2005.01118.x. Hemodial Int. 2005. PMID: 16191054
-
Predicting episodes of hypotension by continuous blood volume monitoring among critically ill patients in acute renal failure on intermittent hemodialysis.Dynamics. 2007 Fall;18(3):19-24. Dynamics. 2007. PMID: 17879763 Clinical Trial.
-
Relative blood volume measurements during hemodialysis: comparisons between three noninvasive devices.Hemodial Int. 2007 Oct;11(4):448-55. doi: 10.1111/j.1542-4758.2007.00216.x. Hemodial Int. 2007. PMID: 17922743 Clinical Trial.
-
Cardiovascular stability on hemodialysis.Adv Nephrol Necker Hosp. 1987;16:351-70. Adv Nephrol Necker Hosp. 1987. PMID: 3101431 Review. No abstract available.
Cited by
-
Design and construct an optical device to determine relative blood volume in patients undergoing hemodialysis.Iran Red Crescent Med J. 2014 Apr;16(4):e15603. doi: 10.5812/ircmj.15603. Epub 2014 Apr 5. Iran Red Crescent Med J. 2014. PMID: 24910804 Free PMC article.
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical
Research Materials