Evaluation of "point of care" devices in the measurement of low blood glucose in neonatal practice
- PMID: 15210675
- PMCID: PMC1721720
- DOI: 10.1136/adc.2003.033548
Evaluation of "point of care" devices in the measurement of low blood glucose in neonatal practice
Abstract
Background: Low blood glucose in newborns is difficult to detect clinically. Hence a reliable "point of care" device (glucometer) for early detection and treatment of low glucose is needed.
Objective: To evaluate the performance of five readily available glucometers for the detection of low blood glucose in newborn infants.
Method: Glucostix measurements were taken for newborns with risk factors using a Reflolux S (Boehringer) glucometer. If the initial reading was low (< 2.6 mmol/l), further measurements were taken with two other glucometers (phase I, Advantage and Glucotrend (Roche); phase II, Elite XL (Bayer) and Precision (Abbott)), and plasma glucose was measured in the laboratory (Aeroset; Abbott).
Results: Over 10 months, 101 specimens were collected from 71 newborns (57 in phase I; 44 in phase II). The Advantage glucometer usually overestimated blood glucose with a mean difference of 1.07 mmol/l (p < 0.01) at all low glucose ranges. The Glucotrend, Precision, and Elite XL glucometers performed better; the mean differences were not significantly different from the laboratory measured value (0.17 mmol/l (p = 0.37); -0.12 mmol/l (p = 0.13), and 0.24 mmol/l (p = 0.13) respectively). For detection of glucose concentrations < 2.6 mmol/l, the Precision glucometer had the highest sensitivity (96.4%) and negative predictive value (90%). For lower glucose concentrations (< 2.0 mmol/l), the Glucotrend glucometer performed even better (sensitivity 92.3%, negative predictive value 96.3%).
Conclusion: Point of care devices should have good precision in the low glucose concentration range, sensitivity, and accuracy for early detection of neonatal hypoglycaemia. None of the five glucometers was satisfactory as the sole measuring device. The Glucotrend and Precision glucometers have the greatest sensitivity and negative predictive value. However, confirmation with laboratory measurements of plasma glucose and clinical assessment are still of the utmost importance.
Figures

Similar articles
-
Accuracy of bedside glucose measurement from three glucometers in critically ill patients.Crit Care Med. 2008 Nov;36(11):3062-6. doi: 10.1097/CCM.0b013e318186ffe6. Crit Care Med. 2008. PMID: 18824915
-
Evaluation of the Glucometer Elite XL device for screening for neonatal hypoglycaemia.Eur J Pediatr. 2005 Nov;164(11):660-4. doi: 10.1007/s00431-005-1733-9. Epub 2005 Jul 22. Eur J Pediatr. 2005. PMID: 16041524
-
Clinical utility and accuracy of a blood glucose meter for the detection of neonatal hypoglycemia.J Coll Physicians Surg Pak. 2004 Apr;14(4):225-8. J Coll Physicians Surg Pak. 2004. PMID: 15228827
-
Measurement of glucose levels in the newborn.Early Hum Dev. 2010 May;86(5):263-7. doi: 10.1016/j.earlhumdev.2010.05.005. Epub 2010 Jun 9. Early Hum Dev. 2010. PMID: 20542649 Review.
-
[Neonatal hypoglycemia].Tidsskr Nor Laegeforen. 2001 May 30;121(14):1713-6. Tidsskr Nor Laegeforen. 2001. PMID: 11446016 Review. Norwegian.
Cited by
-
Screening for hypoglycemia at the bedside in the neonatal intensive care unit (NICU) with the Abbott PCx glucose meter.BMC Pediatr. 2006 Nov 3;6:28. doi: 10.1186/1471-2431-6-28. BMC Pediatr. 2006. PMID: 17083737 Free PMC article.
-
The Accutrend sensor glucose analyzer may not be adequate in bedside testing for neonatal hypoglycemia.Eur J Pediatr. 2006 Feb;165(2):99-103. doi: 10.1007/s00431-005-0013-z. Epub 2005 Oct 19. Eur J Pediatr. 2006. PMID: 16235054
-
Continuous glucose monitoring in preterm infants: evaluation by a modified Clarke error grid.Ital J Pediatr. 2016 Mar 9;42:29. doi: 10.1186/s13052-016-0236-9. Ital J Pediatr. 2016. PMID: 26960676 Free PMC article.
-
Development of a prediction model for neonatal hypoglycemia risk factors: a retrospective study.Front Endocrinol (Lausanne). 2023 Jul 17;14:1199628. doi: 10.3389/fendo.2023.1199628. eCollection 2023. Front Endocrinol (Lausanne). 2023. PMID: 37529595 Free PMC article.
-
Continuous glucose monitoring in newborn infants: how do errors in calibration measurements affect detected hypoglycemia?J Diabetes Sci Technol. 2014 May;8(3):543-50. doi: 10.1177/1932296814524857. Epub 2014 Feb 27. J Diabetes Sci Technol. 2014. PMID: 24876618 Free PMC article.
References
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Medical