Ionescu-Shiley pericardial xenograft valve: Hemodynamic evaluation and early clinical follow-up of 326 patients
- PMID: 15216265
- PMCID: PMC287845
Ionescu-Shiley pericardial xenograft valve: Hemodynamic evaluation and early clinical follow-up of 326 patients
Abstract
Dissatisfaction with the hemodynamic characteristics of available porcine valves prompted a clinical trial of the Ionescu-Shiley percardial xenograft (ISPX) valve. Three hundred fifty-six ISPX valves were implanted consecutively in 326 patients. Operative mortality was 2.6% (2/75) for aortic valve replacement alone and 7.7% (12/155) for aortic valve replacements that included reoperations and combined procedures such as mitral commissurotomy, annuloplasty, and coronary artery bypass. Operative mortality for all patients who underwent mitral valve replacement was 9.5% (14/147). The mean peak systolic gradient pressure in the aortic position was 5.4 mm Hg overall and 4.27 mm Hg with the size 19 mm valve. There were no embolic episodes in patients who received the ISPX valve in the aortic position. The available data indicate that the rate of peripheral embolism with the ISPX valve compares favorably with that of porcine valves. Considering its hemodynamic advantage, if the longterm durability of the full-orifice Ionescu-Shiley pericardial xenograft valve continues to be confirmed by follow-up studies, it is our opinion that it is the biologic valve of choice.
Similar articles
-
Rest and exercise hemodynamics following aortic valve replacement. A comparison between 19 and 21 mm Ionescu-Shiley pericardial and Carpentier-Edwards porcine valves.J Thorac Cardiovasc Surg. 1985 Nov;90(5):750-5. J Thorac Cardiovasc Surg. 1985. PMID: 4058047
-
Clinical results of pericardial xenograft valves: the Ionescu-Shiley and Hancock valves.Can J Cardiol. 1988 Sep;4(6):328-32. Can J Cardiol. 1988. PMID: 3179796
-
Clinical and hemodynamic performance of the Ionescu-Shiley valve in the small aortic root. Results in 117 patients with 17 and 19 mm valves.J Thorac Cardiovasc Surg. 1989 Dec;98(6):1087-95. J Thorac Cardiovasc Surg. 1989. PMID: 2586125
-
Long-term follow-up of the Ionescu-Shiley mitral pericardial xenograft.J Thorac Cardiovasc Surg. 1984 Nov;88(5 Pt 1):758-63. J Thorac Cardiovasc Surg. 1984. PMID: 6492841
-
Long-term results of bioprosthetic mitral valve replacement: the pericardial perspective.J Cardiovasc Surg (Torino). 2004 Oct;45(5):449-54. J Cardiovasc Surg (Torino). 2004. PMID: 15736567 Review.
Cited by
-
Cardiac valve replacement: Experience in 2400 patients.Cardiovasc Dis. 1981 Mar;8(1):7-14. Cardiovasc Dis. 1981. PMID: 15216220 Free PMC article.
-
Noninvasive assessment of the Ionescu-Shiley pericardial xenograft in the mitral position: preliminary experience.Tex Heart Inst J. 1982 Sep;9(3):275-84. Tex Heart Inst J. 1982. PMID: 15226927 Free PMC article.
-
Clinical experience with the Ionescu-Shiley xenograft valve: four to five-year follow-up.Tex Heart Inst J. 1982 Sep;9(3):285-92. Tex Heart Inst J. 1982. PMID: 15226928 Free PMC article.
-
Open-heart surgery at the Texas Heart Institute: the first 50,000 operations.Tex Heart Inst J. 1983 Jun;10(2):107-13. Tex Heart Inst J. 1983. PMID: 15227122 Free PMC article. No abstract available.
-
Echocardiographic assessment of the Ionescu-Shiley pericardial xenograft in the aortic position.Cardiovasc Dis. 1981 Jun;8(2):210-215. Cardiovasc Dis. 1981. PMID: 15216209 Free PMC article.
References
LinkOut - more resources
Full Text Sources