Abstract
Sequential hemodynamic studies consisting of one preoperative and three postoperative cardiac catheterizations were undertaken in 19 patients at approximately 1, 4, and 6 years after valve replacement. Thirteen patients had aortic and six patients had mitral pericardial xenograft valves. Significant improvement was noted in various hemodynamic parameters in both groups of patients at the postoperative studies as compared to the preoperative findings. In patients with aortic valve replacement, the peak systolic gradient was 7.0 +/- 1.2 mm Hg at rest, and 10.0 +/- 1.6 mm Hg after exercise during the third postoperative study. The calculated xenograft orifice areas were 1.6 +/- 0.1 cm(2) at rest and 2.1 +/- 0.2 cm(2) after exercise in the aortic group, and 2.2 +/- 0.2 cm(2) at rest and 2.7 +/- 0.2 cm(2) after exercise in the mitral group at the third postoperative investigation. No significant change was noted among the three postoperative studies in the hemodynamic parameters. The results confirm the maintenance of functional performance of the Ionescu-Shiley pericardial xenograft up to 73 months after valve replacement.
Similar articles
-
Ionescu-Shiley pericardial xenograft valve: Hemodynamic evaluation and early clinical follow-up of 326 patients.Cardiovasc Dis. 1980 Jun;7(2):137-148. Cardiovasc Dis. 1980. PMID: 15216265 Free PMC article.
-
Rest and exercise hemodynamics following aortic valve replacement. A comparison between 19 and 21 mm Ionescu-Shiley pericardial and Carpentier-Edwards porcine valves.J Thorac Cardiovasc Surg. 1985 Nov;90(5):750-5. J Thorac Cardiovasc Surg. 1985. PMID: 4058047
-
Fatigue-induced failure of the Ionescu-Shiley pericardial xenograft in the mitral position. In vivo and in vitro correlation and a proposed classification.J Thorac Cardiovasc Surg. 1984 Jun;87(6):836-44. J Thorac Cardiovasc Surg. 1984. PMID: 6727407
-
[Exercise capacity after heart valve replacement].Z Kardiol. 1994;83 Suppl 3:111-20. Z Kardiol. 1994. PMID: 7941657 Review. German.
-
Long-term results of bioprosthetic mitral valve replacement: the pericardial perspective.J Cardiovasc Surg (Torino). 2004 Oct;45(5):449-54. J Cardiovasc Surg (Torino). 2004. PMID: 15736567 Review.
Cited by
-
Ionescu-Shiley pericardial xenograft valve: Hemodynamic evaluation and early clinical follow-up of 326 patients.Cardiovasc Dis. 1980 Jun;7(2):137-148. Cardiovasc Dis. 1980. PMID: 15216265 Free PMC article.
-
Echocardiographic assessment of the Ionescu-Shiley pericardial xenograft in the aortic position.Cardiovasc Dis. 1981 Jun;8(2):210-215. Cardiovasc Dis. 1981. PMID: 15216209 Free PMC article.
References
LinkOut - more resources
Full Text Sources