Skip to main page content
U.S. flag

An official website of the United States government

Dot gov

The .gov means it’s official.
Federal government websites often end in .gov or .mil. Before sharing sensitive information, make sure you’re on a federal government site.

Https

The site is secure.
The https:// ensures that you are connecting to the official website and that any information you provide is encrypted and transmitted securely.

Access keys NCBI Homepage MyNCBI Homepage Main Content Main Navigation
Comparative Study
. 2004 May;18(5):790-5.
doi: 10.1007/s00464-003-8922-2. Epub 2004 Apr 6.

Dexterity enhancement with robotic surgery

Affiliations
Comparative Study

Dexterity enhancement with robotic surgery

K Moorthy et al. Surg Endosc. 2004 May.

Abstract

Background: The aim of this study was to quantify the extent of dexterity enhancement in robotic surgery as compared to laparoscopic surgery.

Methods: Ten surgeons with varying laparoscopic suturing experience were asked to place three sutures on a suture pad. The sutures were placed laparoscopically, robotically with 2-D vision and robotically with 3-D vision. The da Vinci systems Application Programming Interface (API) was used for positional data. A validated motion analysis system was used for data retrieval for the laparoscopic task. Custom software was developed for data analysis.

Results: Compared to laparoscopic suturing, when the task was undertaken robotically with 2-D vision there was a 20% reduction in the time taken but this was not significant (p = 0.07). There was a 55% reduction in the path traveled by the right hand (p = 0.01) and a 45% reduction in the path traveled by the left hand (p = 0.008). When the task was undertaken robotically with 3-D vision, there was a 40% reduction in the time taken (p = 0.01). There was a 70% reduction in the path traveled by right hand (p = 0.008) and a 55% reduction by the left hand (p = 0.08).

Conclusions: The presence of wristed instrumentation, tremor abolition, and motion scaling enhance dexterity by nearly 50% as compared to laparoscopic surgery. 3-D vision enhances dexterity by a further 10-15%. In addition, the presence of 3-D vision results in a 93% reduction in skills-based errors.

PubMed Disclaimer

References

    1. Surg Endosc. 1999 May;13(5):469-72 - PubMed
    1. Am J Surg. 1993 Apr;165(4):459-65 - PubMed
    1. Surg Laparosc Endosc. 1996 Jun;6(3):191-7 - PubMed
    1. Arch Surg. 2000 Feb;135(2):229-35 - PubMed
    1. Ann Surg. 1996 Nov;224(5):609-20 - PubMed

Publication types

LinkOut - more resources