Skip to main page content
U.S. flag

An official website of the United States government

Dot gov

The .gov means it’s official.
Federal government websites often end in .gov or .mil. Before sharing sensitive information, make sure you’re on a federal government site.

Https

The site is secure.
The https:// ensures that you are connecting to the official website and that any information you provide is encrypted and transmitted securely.

Access keys NCBI Homepage MyNCBI Homepage Main Content Main Navigation
Comparative Study
. 2004 Jun;12(2):195-207.
doi: 10.1109/TNSRE.2004.827223.

Arrays for chronic functional microstimulation of the lumbosacral spinal cord

Affiliations
Comparative Study

Arrays for chronic functional microstimulation of the lumbosacral spinal cord

Douglas McCreery et al. IEEE Trans Neural Syst Rehabil Eng. 2004 Jun.

Abstract

Our objective is to develop neural prostheses based on an array of microelectrodes implanted into the sacral spinal cord, that will allow persons with spinal cord injuries to regain control of their bladder and bowels. For our chronic cat model, we have developed two microelectrode arrays, one type containing nine discrete activated iridium microelectrodes and the second utilizing silicon substrate probes with multiple electrode sites on each probe. Both types can elicit an increase in the pressure within the urinary bladder of more than 40-mm Hg and/or relaxation of the urethral sphincter. A stimulus of 100 microA and 400 micros/ph at 20 Hz (charge-balanced pulses) was required to induce a large increase in bladder pressure or relaxation of the urethral sphincter. We found that 24 h of continuous stimulation with these parameters induced tissue injury (disrupted neuropil, infiltration of inflammatory cells, and loss of neurons close to the tip sites). However, a neural prosthesis that is intended to restore bladder control after spinal cord injury would not operate continuously. Thus, when this stimulus was applied for 24 h, at a 10% duty cycle (1 min of stimulation, then 9 min without stimulation) only minimal histologic changes were observed.