Verbal communication in MR environments: effect of MR system acoustic noise on speech understanding
- PMID: 15220495
- DOI: 10.1148/radiol.2321030955
Verbal communication in MR environments: effect of MR system acoustic noise on speech understanding
Abstract
Purpose: To assess the masking effect of magnetic resonance (MR)-related acoustic noise and the effect of passive hearing protection on speech understanding.
Materials and methods: Acoustic recordings were made at 1.5 T at patient and operator (interventionalist in the MR suite) locations for relevant pulse sequences. In an audiologic laboratory, speech-to-noise ratios (STNRs) were determined, defined as the difference between the absolute sound pressure levels of MR noise and speech. The recorded noise of the MR sequences was played simultaneously with the recorded sentences at various intensities, and 15 healthy volunteers (seven women, eight men; median age, 27 years) repeated these sentences as accurately as possible. The STNR that corresponded with a 50% correct repetition was used as the measure for speech intelligibility. In addition, the effect of passive hearing protection on speech intelligibility was tested by using an earplug model.
Results: Overall, speech understanding was reduced more at operator than at patient location. Most problematic were fast gradient-recalled-echo train and spiral k-space sequences. As the absolute sound pressure level of these sequences was approximately 100 dB at patient location, the vocal effort needed to attain 50% intelligibility was shouting (>77 dB). At operator location, less effort was required because of the lower sound pressure levels of the MR noise. Fast spoiled gradient-recalled-echo and echo-planar imaging sequences showed relatively favorable results with raised voice at operator location and loud speaking at patient location. The use of hearing protection slightly improved STNR.
Conclusion: At 1.5 T, the level of MR noise requires that large vocal effort is used, at the operator and especially at the patient location. Depending on the specific MR sequence used, loud speaking or shouting is needed to achieve adequate bidirectional communication with the patient. The wearing of earplugs improves speech intelligibility.
Copyright RSNA, 2004
Similar articles
-
Speech intelligibility and passive, level-dependent earplugs.Ear Hear. 2011 Sep-Oct;32(5):642-9. doi: 10.1097/AUD.0b013e31821478c8. Ear Hear. 2011. PMID: 21407078 Clinical Trial.
-
Double hearing protection and speech intelligibility-room for improvement.Aviat Space Environ Med. 2001 Apr;72(4):400-4. Aviat Space Environ Med. 2001. PMID: 11318022 Clinical Trial.
-
Speech recognition in noise with active and passive hearing protectors: a comparative study.J Acoust Soc Am. 2011 Jun;129(6):3702-15. doi: 10.1121/1.3575599. J Acoust Soc Am. 2011. PMID: 21682395
-
Noise and communication: a three-year update.Noise Health. 2012 Nov-Dec;14(61):281-6. doi: 10.4103/1463-1741.104894. Noise Health. 2012. PMID: 23257578 Review.
-
A Review of MRI Acoustic Noise Outputs and Hearing Protection Device Performance.J Magn Reson Imaging. 2025 May;61(5):2083-2093. doi: 10.1002/jmri.29665. Epub 2024 Dec 17. J Magn Reson Imaging. 2025. PMID: 39690111 Review.
Cited by
-
Image quality assessment of silent T2 PROPELLER sequence for brain imaging in infants.Br J Radiol. 2018 Feb;91(1083):20170680. doi: 10.1259/bjr.20170680. Epub 2017 Dec 22. Br J Radiol. 2018. PMID: 29212354 Free PMC article.
-
Can hockey playoffs harm your hearing?CMAJ. 2006 Dec 5;175(12):1541-2. doi: 10.1503/cmaj.060789. CMAJ. 2006. PMID: 17146091 Free PMC article. No abstract available.
-
Image quality and subject experience of quiet T1-weighted 7-T brain imaging using a silent gradient coil.Eur Radiol Exp. 2022 Aug 31;6(1):36. doi: 10.1186/s41747-022-00293-x. Eur Radiol Exp. 2022. PMID: 36042139 Free PMC article.
-
Quiet PROPELLER MRI techniques match the quality of conventional PROPELLER brain imaging techniques.AJNR Am J Neuroradiol. 2015 Jun;36(6):1124-7. doi: 10.3174/ajnr.A4235. Epub 2015 Feb 12. AJNR Am J Neuroradiol. 2015. PMID: 25678482 Free PMC article.
-
Reduced acoustic noise in diffusion tensor imaging on a compact MRI system.Magn Reson Med. 2018 Jun;79(6):2902-2911. doi: 10.1002/mrm.26949. Epub 2017 Oct 2. Magn Reson Med. 2018. PMID: 28971512 Free PMC article.
MeSH terms
LinkOut - more resources
Full Text Sources
Medical
Research Materials
Miscellaneous