Patient radiation dose at CT urography and conventional urography
- PMID: 15220498
- DOI: 10.1148/radiol.2321030222
Patient radiation dose at CT urography and conventional urography
Abstract
Purpose: To measure and compare patient radiation dose from computed tomographic (CT) urography and conventional urography and to compare these doses with dose estimates determined from phantom measurements.
Materials and methods: Patient skin doses were determined by placing a thermoluminescent dosimeter (TLD) strip (six TLD chips) on the abdomen of eight patients examined with CT urography and 11 patients examined with conventional urography. CT urography group consisted of two women and six men (mean age, 55.5 years), and conventional urography group consisted of six women and five men (mean age, 58.9 years). CT urography protocol included three volumetric acquisitions of the abdomen and pelvis. Conventional urography protocol consisted of acquisition of several images involving full nephrotomography and oblique projections. Mean and SD of measured patient doses were compared with corresponding calculated doses and with dose measured on a Lucite pelvic-torso phantom. Correlation coefficient (R(2)) was calculated to compare measured and calculated skin doses for conventional urography examination, and two-tailed P value significance test was used to evaluate variation in effective dose with patient size. Radiation risk was calculated from effective dose estimates.
Results: Mean patient skin doses for CT urography measured with TLD strips and calculated from phantom data (CT dose index) were 56.3 mGy +/- 11.5 and 54.6 mGy +/- 4.1, respectively. Mean patient skin doses for conventional urography measured with TLD strips and calculated as entrance skin dose were 151 mGy +/- 90 and 145 mGy +/- 76, respectively. Correlation coefficient between measured and calculated skin doses for conventional urography examinations was 0.95. Mean effective dose estimates for CT urography and conventional urography were 14.8 mSv +/- 90.0 and 9.7 mSv +/- 3.0, respectively. Mean effective doses estimated for the pelvic-torso phantom were 15.9 mSv (CT urography) and 7.8 mSv (conventional urography).
Conclusion: Standard protocol for CT urography led to higher mean effective dose, approximately 1.5 times the radiation risk for conventional urography. Patient dose estimates should be taken into consideration when imaging protocols are established for CT urography.
Copyright RSNA, 2004
Comment in
-
Patient radiation dose at CT urography and conventional urography.J Urol. 2005 Mar;173(3):923-4. doi: 10.1097/01.ju.0000151885.61744.1d. J Urol. 2005. PMID: 15711330 No abstract available.
Similar articles
-
Radiation dose reduction in thoracic and abdomen-pelvic CT using tube current modulation: a phantom study.J Appl Clin Med Phys. 2014 Jan 8;16(1):5135. doi: 10.1120/jacmp.v16i1.5135. J Appl Clin Med Phys. 2014. PMID: 25679153 Free PMC article.
-
Verification of organ doses calculated by a dose monitoring software tool based on Monte Carlo Simulation in thoracic CT protocols.Acta Radiol. 2018 Mar;59(3):322-326. doi: 10.1177/0284185117716199. Epub 2017 Jun 15. Acta Radiol. 2018. PMID: 28618854
-
Skin and thyroid dosimetry in cervical spine screening: two methods for evaluation and a comparison between a helical CT and radiographic trauma series.AJR Am J Roentgenol. 2002 Oct;179(4):933-7. doi: 10.2214/ajr.179.4.1790933. AJR Am J Roentgenol. 2002. PMID: 12239041
-
Studying the dose level for different X-ray energy conventional radiography by TLD-100.Appl Radiat Isot. 2022 Mar;181:110066. doi: 10.1016/j.apradiso.2021.110066. Epub 2021 Dec 18. Appl Radiat Isot. 2022. PMID: 34968881 Review.
-
Radiation protection issues in dynamic contrast-enhanced (perfusion) computed tomography.Eur J Radiol. 2015 Dec;84(12):2347-58. doi: 10.1016/j.ejrad.2014.11.011. Epub 2014 Nov 20. Eur J Radiol. 2015. PMID: 25480677 Review.
Cited by
-
Lower-dose CT urography (CTU) with iterative reconstruction technique in children - initial experience and examination protocol.Pol J Radiol. 2014 Jun 8;79:137-44. doi: 10.12659/PJR.890729. eCollection 2014. Pol J Radiol. 2014. PMID: 24920965 Free PMC article.
-
Stratifying risk of disease in haematuria patients using machine learning techniques to improve diagnostics.Front Oncol. 2024 May 8;14:1401071. doi: 10.3389/fonc.2024.1401071. eCollection 2024. Front Oncol. 2024. PMID: 38779086 Free PMC article.
-
Use of the bolus tracking technique for the tomographic evaluation of the uretero-vesicular junction in dogs and assessment of dose records.BMC Vet Res. 2016 Mar 29;12:64. doi: 10.1186/s12917-016-0690-z. BMC Vet Res. 2016. PMID: 27026013 Free PMC article.
-
Dose reduction in multidetector CT of the urinary tract. Studies in a phantom model.Eur Radiol. 2006 Sep;16(9):1982-9. doi: 10.1007/s00330-005-0138-5. Epub 2006 Mar 28. Eur Radiol. 2006. PMID: 16568267
-
Multidetector CT urography: experimental analysis of radiation dose reduction in an animal model.Eur Radiol. 2007 Sep;17(9):2318-24. doi: 10.1007/s00330-006-0565-y. Epub 2007 Feb 14. Eur Radiol. 2007. PMID: 17429649
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Medical