Performance evaluation of the positron scanner ECAT EXACT
- PMID: 1522276
- DOI: 10.1097/00004728-199209000-00024
Performance evaluation of the positron scanner ECAT EXACT
Abstract
The Cologne Special is a prototype of the ECAT EXACT (model 921), a new generation of Siemens-CTI PET scanners. It consists of three rings of 48 BGO block detectors each, covering an axial field of view of 16.2 cm with a patient port of 56.2 cm diameter. This results in a total of 24 rings with 384 crystals each, giving 47 contiguous image planes in two-dimensional (2D) mode. Total system sensitivity is 216 kcps/microCi/ml for a 20 cm cylinder phantom in 2D. This increases to 1.5 Mcps/microCi/ml in 3D. Data are acquired in the stationary mode only (no wobble motion), resulting in a transaxial spatial resolution of better than 6 mm full width at half-maximum (FWHM) at the center, which degrades to 7.5 mm tangentially and 9.6 mm radially at a radius of 20 cm. Average axial resolution changes from 5.0 mm FWHM at the center to 8.1 mm at R = 20 cm. Count rate performance was investigated at different low energy discriminator settings and found to be linear up to 2.5 microCi/ml with a 20 cm phantom. The magnitude and distribution of scatter were evaluated for both septa-extended and septa-retracted conditions for a range of energy thresholds. Brain, heart, and whole-body studies with the new tomograph demonstrate the versatility of its applications without compromising on physical performance.
Similar articles
-
The ECAT EXACT HR: performance of a new high resolution positron scanner.J Comput Assist Tomogr. 1994 Jan-Feb;18(1):110-8. J Comput Assist Tomogr. 1994. PMID: 8282858
-
Performance evaluation of a large axial field-of-view PET scanner: SET-2400W.Ann Nucl Med. 1997 Nov;11(4):307-13. doi: 10.1007/BF03165298. Ann Nucl Med. 1997. PMID: 9460522
-
Performance evaluation of a whole-body PET scanner using the NEMA protocol. National Electrical Manufacturers Association.J Nucl Med. 1997 Oct;38(10):1614-23. J Nucl Med. 1997. PMID: 9379202
-
PET instrumentation: what are the limits?Semin Nucl Med. 1998 Jul;28(3):247-67. doi: 10.1016/s0001-2998(98)80030-5. Semin Nucl Med. 1998. PMID: 9704366 Review.
-
Instrumentation for positron emission tomography: tomographs and data processing and display systems.Semin Nucl Med. 1992 Jul;22(3):162-81. doi: 10.1016/s0001-2998(05)80144-8. Semin Nucl Med. 1992. PMID: 1523414 Review.
Cited by
-
Bismuth germanate coupled to near ultraviolet silicon photomultipliers for time-of-flight PET.Phys Med Biol. 2016 Sep 21;61(18):L38-L47. doi: 10.1088/0031-9155/61/18/L38. Epub 2016 Sep 2. Phys Med Biol. 2016. PMID: 27589153 Free PMC article.
-
Correlation of 18F-FLT and 18F-FDG uptake on PET with Ki-67 immunohistochemistry in non-small cell lung cancer.Eur J Nucl Med Mol Imaging. 2007 Oct;34(10):1610-6. doi: 10.1007/s00259-007-0449-7. Epub 2007 May 26. Eur J Nucl Med Mol Imaging. 2007. PMID: 17530250
-
Human dopamine receptor D2/D3 availability predicts amygdala reactivity to unpleasant stimuli.Hum Brain Mapp. 2010 May;31(5):716-26. doi: 10.1002/hbm.20900. Hum Brain Mapp. 2010. PMID: 19904802 Free PMC article.
-
From PET detectors to PET scanners.Eur J Nucl Med Mol Imaging. 2003 Nov;30(11):1574-97. doi: 10.1007/s00259-003-1266-2. Epub 2003 Oct 2. Eur J Nucl Med Mol Imaging. 2003. PMID: 14579100 Review.
-
Comparison of different SUV-based methods for monitoring cytotoxic therapy with FDG PET.Eur J Nucl Med Mol Imaging. 2004 Nov;31(11):1471-8. doi: 10.1007/s00259-004-1626-6. Epub 2004 Jul 15. Eur J Nucl Med Mol Imaging. 2004. PMID: 15257418
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources